Hardware Reference
5HDJHQWJDVIORZFRQWUROPRGXOH
17
•
The flow control display shows the gas flow as a percentage of the total possible
flow (5 ml/min for
methane
).
•
The flow control knob adjusts the gas flow. If the selected flow rate can not be
achieved or maintained, the numbers in the flow control display will flash.
•
The CI calibration valve is controlled by the ChemStation software, and opens
automatically during CI autotune or manual tuning, allowing PFDTD to diffuse
into the ion source.
•
The isolation valve prevents contamination of the flow control module by
atmosphere while the MSD is vented or by PFTBA during EI operation.
Reagent gas select valve
Mass flow controller
Rear flow module cover
Flow control board (under
Isolation valve
CI calibration valve
Reagent gas inlet lines
Control panel
EI/CI GC/MSD interface
& $ 8 7 , 2 1
Try to avoid tuning any more often than absolutely necessary to minimize PFDTD
background and ion source contamination.
Содержание 5973 MSD
Страница 1: ...H Hardware Manual HP 5973 MSD PCI NCI ...
Страница 6: ...6 ...
Страница 7: ...1 General Information Specifications 9 Reagent gases and regulators 10 ...
Страница 13: ... Hardware Reference 13 ...
Страница 25: ... Chemical Ionization Theory 3RVLWLYH WKHRU 25 ...
Страница 31: ... Chemical Ionization Theory 1HJDWLYH WKHRU 31 ...
Страница 34: ...34 ...
Страница 92: ...92 ...
Страница 103: ...8 Parts GC MSD interface parts 106 CI ion source parts 108 Flow control module parts 110 Miscellaneous parts 113 ...
Страница 114: ...114 ...
Страница 120: ...120 ...
Страница 122: ...H Manual Part Number G1999 90001 Copyright 1998 Hewlett Packard Printed in USA 2 98 ...