Hardware Reference
&,LRQVRXUFH
CI ion source
The CI ion source is similar to the EI source, but only has one part in common with
the EI source — the entrance lens. The single CI filament has a straight wire and a
reflector. There is a “dummy” filament to provide connections for the other wires.
The holes in the ion source (electron-entrance and ion-exit) are very small (0.5
mm), making it possible to pressurize the ionization chamber. Both the source body
and the plate are at repeller potential, electrically isolated from the radiator and the
CI interface tip. The seal for the interface tip ensures a leak-tight seal and electrical
isolation between the CI interface and ion source. Large slots in the drawout
cylinder and ion-source body provide enhanced pumpout for the ion source optics.
CI ion source body
Source heater and
sensor cables
CI interface tip
CI interface tip seal
CI source heater
assembly
Ion focus lens
CI interface cover
CI filament
Entrance lens
Содержание 5973 MSD
Страница 1: ...H Hardware Manual HP 5973 MSD PCI NCI ...
Страница 6: ...6 ...
Страница 7: ...1 General Information Specifications 9 Reagent gases and regulators 10 ...
Страница 13: ... Hardware Reference 13 ...
Страница 25: ... Chemical Ionization Theory 3RVLWLYH WKHRU 25 ...
Страница 31: ... Chemical Ionization Theory 1HJDWLYH WKHRU 31 ...
Страница 34: ...34 ...
Страница 92: ...92 ...
Страница 103: ...8 Parts GC MSD interface parts 106 CI ion source parts 108 Flow control module parts 110 Miscellaneous parts 113 ...
Страница 114: ...114 ...
Страница 120: ...120 ...
Страница 122: ...H Manual Part Number G1999 90001 Copyright 1998 Hewlett Packard Printed in USA 2 98 ...