![HP 54620A Скачать руководство пользователя страница 162](http://html.mh-extra.com/html/hp/54620a/54620a_users-and-service-manual_161475162.webp)
2
Measure the power supply voltages at L1, L2, and L3 on the system
board.
Make sure that the voltage measurements are within the following tolerances.
+
5.1 V
±
150 mV (
+
4.95 V to
+
5.25 V)
+
15.75 V
±
787 mV (
+
14.96 V to
+
16.54 V)
−
15.75 V
±
787 mV (
−
14.96 V to
−
16.54 V)
If the
+
5.1 V measurement is out of tolerance, adjust the
+
5.1 V adjustment
on the power supply. The
±
15.75 V supplies are not adjustable and are
dependent upon the
+
5.1 V supply. If adjusting the power supply does not
bring all the voltages within tolerance, see “Troubleshooting the Logic
Analyzer” in this chapter.
Power Supply Adjustment
Figure 54
Testing, Adjusting, and Troubleshooting the Analyzer
To adjust the power supply
161
Содержание 54620A
Страница 7: ...6 ...
Страница 13: ...12 ...
Страница 17: ...16 ...
Страница 18: ...1 Getting Started ...
Страница 20: ...Using the Logic Analyzer Figure 1 Getting Started 19 ...
Страница 52: ...2 Making Analyzer Measurements ...
Страница 121: ...120 ...
Страница 122: ...3 Solving Problems ...
Страница 127: ...126 ...
Страница 128: ...4 Ensuring Accurate Measurements ...
Страница 150: ...5 Testing Adjusting and Troubleshooting the Analyzer ...
Страница 180: ...6 Replaceable Parts ...
Страница 193: ...Exploded View of Logic Analyzer Figure 60 Replaceable Parts To order a replacement part 192 ...
Страница 196: ...7 Performance Characteristics ...
Страница 208: ...8 Messages ...
Страница 217: ...216 ...
Страница 239: ......
Страница 241: ...240 ...