![HP 54620A Скачать руководство пользователя страница 137](http://html.mh-extra.com/html/hp/54620a/54620a_users-and-service-manual_161475137.webp)
The range of available delay is different with negative delay values. Suppose
you set the time reference to Right, at a sweep speed of 5 ms, with a delay
value of -80
µ
s, and you perform a single acquisition. The display would look
like the following:
Single Acquisition, Time Reference Right, Negative Delay of 80
µ
s
The time reference
is set to Right
The trigger point is
off-screen to the right
due to negative delay
Figure 40
Ensuring Accurate Measurements
Time base and Acquisition