Chapter 4 Command Reference
:STATus Subsystem
4-34
Programming Guide
4
COMMAND
:STATus:OPERation:NTRansition <non-decimal numeric> | <NRf>
Sets or queries the negative transition filter for the Operation status
reporting structure.
RANGE
The range of the <non-decimal numeric> or <NRf> parameter is
0 to 32,767.
QUERY RESPONSE
Numeric data transferred as ASCII bytes in <NR1> format.
COMMENTS
•
The parameter and the query response value, when rounded to an
integer value and expressed in base 2 (binary), represent the bit
values of the negative transition filter.
•
The value of unused bits is zero when queried and is ignored when set.
•
A TRUE bit in the negative transition filter specifies that a negative
(TRUE to FALSE) transition of the corresponding bit in the Operation
Condition Status Register generates the corresponding event in the
Operation Event Status Register.
•
At power-on and STAT:PRES, the negative transition filter is preset
such that each bit is a 0 (FALSE).
•
This value is unaffected by *RST, *CLS, and *SAV/*RCL.
RELATED
FRONT-PANEL
KEYS
None
COMMAND
:STATus:OPERation:PTRansition . . . <non-decimal numeric> | <NRf>
Sets or queries the positive transition filter for the Operation status
reporting structure.
RANGE
The range of the <non-decimal numeric> or <NRf> parameters is
0 to 32,767.
QUERY RESPONSE
Numeric data transferred as ASCII bytes in <NR1> format.
COMMENTS
•
The parameter and the query response value, when rounded to an
integer value and expressed in base 2 (binary), represent the bit
values of the positive transition filter.
Содержание 53150A Series
Страница 1: ... Programming Guide HP 53150A 151A 152A Microwave Frequency Counter ...
Страница 2: ......
Страница 10: ...Contents x Programming Guide ...
Страница 11: ...1 Before You Start ...
Страница 20: ...Chapter 1 Before You Start Related Documentation 1 10 Programming Guide 1 ...
Страница 21: ...2 Command Summary A Quick Reference ...
Страница 41: ...3 Programming Your Counter for Remote Operation ...
Страница 100: ...Chapter 3 Programming Your Counter for Remote Operation Programming Examples 3 60 Programming Guide 3 ...
Страница 101: ...4 Command Reference A Dictionary ...
Страница 162: ...Chapter 4 Command Reference Common Commands 4 62 Programming Guide 4 ...
Страница 163: ...5 Errors ...
Страница 174: ...Chapter 5 Errors Error Types 5 12 Programming Guide 5 ...
Страница 184: ......
Страница 186: ... Manual Part Number 53150 90002 Printed in U S A NOVEMBER 1997 ...