Chapter 4 Command Reference
:STATus Subsystem
Programming Guide
4-33
4
COMMAND
:STATus:OPERation[:EVENt]?
Queries the status of the Operation Event Status Register.
QUERY RESPONSE
•
Numeric data transferred as ASCII bytes in <NR1> format.
•
Range is 0 to 32,767.
•
The query response value is an integer formed by the binary
weighting of bits. The value of unused bits is zero.
COMMENTS
•
Each event bit in the Operation Event Status Register corresponds to
a specific condition bit in the Operation Condition Status Register;
this allows the Operation Event Status Register to detect changes in
conditions.
•
An event becomes TRUE when the associated condition makes the
transition specified by the transition filters.
•
The event bits, once set, are “sticky”—i.e., they cannot be cleared until
they are read, even if they do not reflect the current status of a related
condition.
•
The Operation Event Status Register is cleared by *CLS, by
:STAT:OPER[:EVEN]?, and at power-on.
RELATED
FRONT-PANEL
KEYS
None
Содержание 53150A Series
Страница 1: ... Programming Guide HP 53150A 151A 152A Microwave Frequency Counter ...
Страница 2: ......
Страница 10: ...Contents x Programming Guide ...
Страница 11: ...1 Before You Start ...
Страница 20: ...Chapter 1 Before You Start Related Documentation 1 10 Programming Guide 1 ...
Страница 21: ...2 Command Summary A Quick Reference ...
Страница 41: ...3 Programming Your Counter for Remote Operation ...
Страница 100: ...Chapter 3 Programming Your Counter for Remote Operation Programming Examples 3 60 Programming Guide 3 ...
Страница 101: ...4 Command Reference A Dictionary ...
Страница 162: ...Chapter 4 Command Reference Common Commands 4 62 Programming Guide 4 ...
Страница 163: ...5 Errors ...
Страница 174: ...Chapter 5 Errors Error Types 5 12 Programming Guide 5 ...
Страница 184: ......
Страница 186: ... Manual Part Number 53150 90002 Printed in U S A NOVEMBER 1997 ...