Hospira GemStar System Operating Manual
41
Chapter 2. Basic Operation
Priming Pump Sets WITH a Drip Chamber
1
Push down on the flow stop
to
close the cassette.
2
Insert the piercing pin into the
container. Suspend the container
on an I.V. Pole.
3
Squeeze the drip chamber until
filled halfway.
4
Push down on the flow stop
to
open (priming position) the
cassette.
5
Ensure the CAIR (roller) clamp is
open prior to inserting cassette.
6
Invert the cassette approximately
45°.
WARNING: Failure to use the pressure-activated anti-siphon valve
may result in unrestricted flow if the cassette is not
properly installed in the pump.
When using a vented chamber, ALWAYS use an
air-eliminating filter OR set the air sensitivity to
either ON or 2 mL.
Содержание GemStar Series
Страница 9: ...Notes viii Hospira GemStar System Operating Manual ...
Страница 11: ...Chapter 1 Introduction 2 Hospira GemStar System Operating Manual GemStar Pump Layout Shown approximate size ...
Страница 27: ...Notes 18 Hospira GemStar System Operating Manual ...
Страница 67: ...Notes 58 Hospira GemStar System Operating Manual ...
Страница 77: ...Notes 68 Hospira GemStar System Operating Manual ...
Страница 87: ...Notes 78 Hospira GemStar System Operating Manual ...
Страница 95: ...Notes 86 Hospira GemStar System Operating Manual ...
Страница 103: ...Notes 94 Hospira GemStar System Operating Manual ...
Страница 165: ...Notes 156 Hospira GemStar System Operating Manual ...
Страница 173: ...Chapter 14 Operation Test 164 Hospira GemStar System Operating Manual ...
Страница 175: ...Notes 166 Hospira GemStar System Operating Manual ...
Страница 187: ...Chapter 15 Optional System Components 178 Hospira GemStar System Operating Manual A B D C I H G F E J N M K L ...
Страница 193: ...Notes 184 Hospira GemStar System Operating Manual ...
Страница 221: ...Notes 212 Hospira GemStar System Operating Manual ...