Hospira GemStar System Operating Manual
115
Chapter 10. Options Menu
Bolus History
To display the bolus history:
1
Press [
OPTIONS
] and select
HISTORIES.
2
Select
BOLUS HISTORY
from the
Histories Menu.
3
Press [
BACK-UP
] when done to exit
OPTIONS
.
Operation Test
The Operation Test checks several pump functions to ensure the GemStar Pump is
working properly. Refer to Chapter 14 Operation Test on page 157 for instructions.
d
Note: Bolus History is only available if the current
program has a bolus programmed, or had a bolus
programmed at one time. The Bolus History is
maintained in hourly increments for the latest 48 hours.
The Bolus History is cleared when a new program is
entered, a Speed Protocol is recalled, or the History is
cleared.
TIME IS 09:15 AM
FRI, NOV 10, 06
BOL DEL 30
BOL DEM 37
BOL AMT 30.0 mL
06:10A TO 09:14A
Sample Bolus History
Содержание GemStar Series
Страница 9: ...Notes viii Hospira GemStar System Operating Manual ...
Страница 11: ...Chapter 1 Introduction 2 Hospira GemStar System Operating Manual GemStar Pump Layout Shown approximate size ...
Страница 27: ...Notes 18 Hospira GemStar System Operating Manual ...
Страница 67: ...Notes 58 Hospira GemStar System Operating Manual ...
Страница 77: ...Notes 68 Hospira GemStar System Operating Manual ...
Страница 87: ...Notes 78 Hospira GemStar System Operating Manual ...
Страница 95: ...Notes 86 Hospira GemStar System Operating Manual ...
Страница 103: ...Notes 94 Hospira GemStar System Operating Manual ...
Страница 165: ...Notes 156 Hospira GemStar System Operating Manual ...
Страница 173: ...Chapter 14 Operation Test 164 Hospira GemStar System Operating Manual ...
Страница 175: ...Notes 166 Hospira GemStar System Operating Manual ...
Страница 187: ...Chapter 15 Optional System Components 178 Hospira GemStar System Operating Manual A B D C I H G F E J N M K L ...
Страница 193: ...Notes 184 Hospira GemStar System Operating Manual ...
Страница 221: ...Notes 212 Hospira GemStar System Operating Manual ...