Measurement
55
Fig. 56 Oil content extraction from part (immersion)
Note
Do not shake the container with the solvent and with closed lid hard. If it is shaken hard, the
internal pressure will increase due to vaporization of the solvent and the solvent may spray out
from gaps under the lid.
After shaking, open the lid briefly to lower the internal pressure and then promptly close the lid.
Do not apply the ultrasonic cleaner longer than 3 minutes.
5.
Filter the liquid through the filter paper to remove foreign matter (particles, peeled
fragments, etc.), and use it as the measurement liquid.
Zero calibration
Perform zero calibration using the conditions in the table below.
Note
Do not introduce water during zero calibration.
Reference
"Calibration Setting screen" (page 80)
"Zero liquid for calibration" (page 21)
Parameter
Condition
Extraction Time
0 sec
Separation Time
0 sec
Number of Purge
3 or more
Zero liquid for calibration
Use 24 mL of zero liquid for calibration.