Reference
113
Solvent S-316
Characteristics
Solvent S-316 has the characteristics shown below, and satisfies the conditions required of a
solvent for oil content extraction.
Although there is absorption near the 3000 cm
-1
absorption wavelength of the
hydrocarbon group, this absorption can be clearly distinguished from that of oil.
Available for measurement in wide range of temperature due to the boiling point of 134°C
and the melting point of
−
143°C.
Chemically stable in acid, alkali, oil and water.
Little solubility in water.
Involatile, with a low vapor pressure.
Non-flammable, no danger of explosion
Low toxicity for the human body, very safe.
Properties of S-316
Reclamation of solvent
To reduce running costs and help protect the global environment, it is recommended that S-
316 is reclaimed. To reclaim solvent, use the optional SR-305 solvent reclamation unit.
Tip
Solvent reclamation unit (SR-305) uses activated carbon and alumina layers to efficiently reclaim the
solvent.
For the solvent reclamation procedure, refer to the manual for the SR-305 solvent reclamation
unit.
This section describes handling methods and cautionary points for efficient reclamation of
solvent.
Absorbents
The following 2 types of absorbents are used in the solvent reclamation unit.
Activated carbon:
Removes oils, fats, and other substances that do not dissolve in water.
Activated alumina:
Removes water content and substances that easily dissolve in water.
Note
Store absorbents in a dry location. If an absorbent becomes damp, its performance will drop
noticeably.
Chemical formula
Cl(CF
2
-CFCl)
2
Cl
Molecular weight
304
Boiling point
134°C
Melting point
−
143°C
Density
1.75 g/mL (25°C)
Vapor pressure
0.0015 MPa (25°C)
Saturated solubility in water
0.0048 g/100 g (25°C)
Acute oral toxicity (LD50)
52.5 g/kg or more