Maintenance
102
17.
After liquid drainage is finished, press the ESC button.
The start state of the manual measurement mode returns.
18.
Close the right cover.
Drying the measurement cell
If the measurement cell gets wet, it may cause incorrect measurement results.
Maintenance interval guideline
If the measurement cell gets wet
Items required
Tissue paper
Dry air
Measuring cylinder (20 mL) or measuring syringe (20 mL, optional): 1 (for solvent)
Clean solvent S-316 (optional)
Drying procedure
Note
Observe the precautions below when working.
Do not pull out the filter block with the joint attached. The joint may come off or be bent.
Install the water filter correctly. If poorly installed, water content in the solvent may enter the
measurement cell and an abnormal measured value alarm will occur.
During installing the filter block after replacing the water filter, attach the joint firmly to prevent
liquid leakage.
CAUTION
Chemical hazard (solvent S-316)
Inhalation or accidental ingestion of a large amount of solvent S-316 may be harmful.
Observe the following rules when handling:
• Ventilate the work area sufficiently.
•
• Wear a protective mask and protective gloves.
• Wash hands well after handling the solvent.
Take care not to pinch your fingers when opening or closing the right cover.
During closing the right cover, do not release your hand until you hear a click sound.