Maintenance
91
Washing the fan filter
If the filter, attached to the fan vent, starts to clog and the internal temperature rises, accurate
measurement values can no longer be obtained and there is a risk of product failure. Wash the
fan filter on a regular basis.
Maintenance interval guideline
Once a month
Items required
Flathead screwdriver or similar tool
Work procedure
1.
Turn OFF the power.
2.
Use the flathead screwdriver to remove the retainer from the fan vent on the back of
the main unit.
Fig. 102 Removing the fan filter
3.
Remove the fan filter and wash with water to remove dirt.
4.
Allow the fan filter to dry completely.
5.
Re-attach the fan filter and retainer.