![horiba LO-300 Скачать руководство пользователя страница 54](http://html1.mh-extra.com/html/horiba/lo-300/lo-300_instruction-manual_2162291054.webp)
Operation
40
Upper/lower limit level setting
This screen enables to set the threshold of upper/lower limit level.
Fig. 34 Upper/lower limit level setting screen
Tip
Ensure to set the upper/lower limit level as follows: Lower level
Detection level
Upper level.
Also, when lower limit error judgment setting is “OFF”, please set the lower limit level to 4 mA.
Normal detection level may not be output.
Name
Description
Upper level
Any value exceeding this value will be judged as upper limit error level.
Setting range: 4 mA to 24 mA; touch
/
keys to change the setting.
Lower level
Any value falling below this value will be judged as lower limit error level (If Lower error
judge setting is ON).
Setting range:4 mA to 24 mA; touch
/
keys to change the setting.
Upper level
Lower level