Operation
38
Contact output setting
This screen enables to set the contact output mode and the contact output hold time.
Fig. 31 Contact output setting screen
l
Lower limit error judgment setting
This screen enables to set ON/OFF of the lower limit error judgment.
Fig. 32 Lower limit error judgment setting screen
Name
Description
Contact output mode
To change the mode, touch the WARNING button or the ALARM button. When WARNING
is selected, oil film contact is output by oil film warning or oil film alarm. When ALARM is
selected, oil film contact is output only by oil film alarm. For detailed information of contact
output, refer to “ Contact output ” (page 6).
When the optional extended module is incorporated, both “WARNING” and “ALARM”
become the selected state.
Contact output hold time
The setting range is 0 seconds to 99 seconds or
∞
(infinity).
After the output of oil film contact by oil film detection, if there is no oil film on the detection
area any more and the system returns to the normal condition, the oil film contact output
continues for a set period of time. If this time is set to
∞
(infinity), the output continues until
it is manually released.
Contact output mode
Contact output
hold time
Name
Description
Lower limit error
judgment ON/OFF
To switch the lower limit error judgment, touch ON or OFF on the screen.
When ON is selected, lower limit error judgment is performed.
When OFF is selected, lower limit error judgment is not performed.
Lower limit
error judgment
ON/OFF