Measurement setting
56
■
Measurement setting
●
Displaying the latest calibration and inspection data
• In the case of displaying the latest pH calibration data and repeatability
inspection data
You can display the latest pH calibration data and repeatability inspection data.
1. Press the
key to enter the setting mode when "pH" icon is shown on the
main screen.
2. Press the
keys to select "pH" and then press the
key.
3. Press the
keys to select "CAL DATA" (display calibration data) and then
press the
key.
The electrode status based on the calibration result is displayed. You can change the
display item among "OFS" (asymmetry potential) to SLP (sensitivity) to REP
(repeatability) by pressing the
keys.
key
key