
Dissolved oxygen measurement
48
Note
• The criteria of stability judgment in the automatic hold measurement are as follows.
Dissolved oxygen:
Display value change for 10 seconds is less than 0.03
mg/L and temperature change is less than 2.0°C.
Oxygen concentration:
Display value change for 10 seconds is less than
0.1% and temperature change is less than 2.0°C.
Saturated oxygen concentration:
Display value change for 10 seconds is less than
0.3% and temperature change is less than 2.0°C.
• "Ur" (Under) lights when the measured value is below the display range and "Or" (Over)
when the measured value is above the display range. For details on how to correct them,
refer to "The measured value is out of the measurement range" (page 92).
●
Saving the measured value
To save the measurement data, press the
key in the screen that you want to save.
For details, refer to "Saving measurement data in the internal memory" (page 14).
Содержание LAQUAact-PD110
Страница 1: ...Handheld pH ORP Dissolved Oxygen Meter LAQUAact PD110 Instruction Manual CODE GZ0000427800...
Страница 2: ......
Страница 12: ...X M E M O...
Страница 100: ...M E M O 84...
Страница 110: ...M E M O 94...
Страница 117: ...2 Miyanohigashi Kisshoin Minami ku Kyoto 601 8510 Japan http www horiba com...