Troubleshooting
78
●
The indicated value does not change/No response
●
The measured value is outside the display range
When the measured value is below the display range, "Ur" appears. When the measured
value is over the display range, "Or" appears.
●
Repeatability of the measured value is poor
Cause
How to solve problem
The conductivity cell is broken.
Replace the conductivity cell.
The conductivity cell is not
connected correctly.
Connect the conductivity cell correctly.
The instrument is in HOLD state. Cancel the HOLD state.
Instrument defect
Consult your dealer.
Cause
How to solve problem
Sample is out of the
measurement range.
Use a sample within the measurement range.
The conductivity cell cable is
broken.
Replace the conductivity cell.
Cell constant is not set.
Set the cell constant.
Calibration is not performed or
performed incorrectly.
Perform calibration correctly.
Instrument defect
Check as explained below.
Cause
How to solve problem
Effect of the sample solution
Repeatability becomes poor when the conductivity of the
sample changes over time.
The conductivity cell is dirty.
Clean the conductivity cell.
The conductivity cell is broken.
Replace the conductivity cell.