pH measurement
30
Note
・
The criteria of stability judgment in the automatic hold measurement are as follows.
Potential variation for 10 seconds is less than 1 mV (0.015 pH equivalent) and
temperature variation is less than 2.0°C
・
If the measured value is below the display range, "Ur" (under) appears. If the measured
value is above the display range, "Or" (over) appears. For details on the action to take,
refer to “ The measured value is outside the display range ” (page 90).
●
Saving measured values
To save the measurement data, press the
key in the screen that you want to save.
For details, refer to “ Saving measurement data in the internal memory ” (page 11).
●
Printing measured values in the measurement mode
To print the measurement data, press the key in the screen that you want to print. For
details, refer to “ Printing measured values and calibration data ” (page 71).