Maintenance and Storage
66
● Maintenance and Storage of Electrodes
This section describes an overview of the procedures for maintenance and storage of pH, ORP,
ion and conductivity electrodes. For the detailed procedures, refer to the instruction manual for
each electrode.
● How to clean the electrodes
When the tip of an electrode (responsive membrane and liquid junction) becomes dirty, the
response time may slow or an error may occur in the measurement results. To avoid such
errors, clean the electrode. For dirt that cannot be washed off by pure water (or deionized
water), use the cleaning solution indicated below depending on the type of dirt. After cleaning,
rinse the electrode with pure water (or deionized water).
However for pH and ORP electrodes, different cleaning solutions should be used to clean
different types of dirt.
For pH Electrode
For ORP Electrode
Type of dirt
Cleaning solution
General
Diluted neutral cleaning solution
Oil
Alcohol, or diluted neutral cleaning solution
Inorganic
substance
1 mol/L HCl or electrode cleaning solution (model 220)
Protein
Cleaning solution including protein-removing enzyme (model 250)
Alkaline
Dip in 1 mol/L HCl or electrode cleaning solution (model 220) for 1 h to
2 h
Type of dirt
Cleaning solution
General
Dilute neutral cleaning solution (General dishwashing liquid
works reasonably well.)
Oil
Inorganic substance
Immerse dilute nitric acid (1:1 nitric acid)
Содержание LAQUA 2000 Series
Страница 2: ......
Страница 4: ...II MEMO...
Страница 13: ...For Your Safety XI JPN EN61326 1 EN61010 1 II 2...
Страница 54: ...Ion Setup 36 key key key key...
Страница 96: ...Printout Summary 78 Appendix 2 pH mV Relative mV Printer Format Measurement...
Страница 97: ...Printout Summary 79 Ion Conductivity Resistivity...
Страница 98: ...Printout Summary 80 Salinity TDS...
Страница 99: ...Printout Summary 81 Printer Format Data Log...
Страница 100: ...Printout Summary 82 Ion pH Printer Format Calibration...
Страница 101: ...Printout Summary 83 Conductivity Salinity...
Страница 106: ...88...
Страница 107: ......