Conductivity Setup
41
● P1.3 Auto Calibration Setup
1. Press the
key,
P1 COND
screen appears.
2. Press the
ENT
key,
P1.1 CELL
screen appears.
3. Press the
key,
P1.2 UNIT
screen appears.
4. Press the
key,
P1.3 A.CAL
appears.
5. Press the
ENT
key, default
A.CAL ON
screen appears.
6. Use the
keys to change to
OFF
. This enables the manual calibration method.
7. Press the
ENT
key,
P1.3 A.CAL
screen appears. This indicates completion of auto
calibration setup.
key
key
key
key
key
key