Measurement
47
External input 2 (cleaning/hold)
Selection of
command by
contact
(cleaning/hold
command)
When external input 2 is a cleaning command (“tyPE” of Cleaning
operation setting is set to “or”, “And”, or “trg”), if this contact is closed,
automatic cleaning and its associated operations will be started.
Note:
The following display cannot be changed.
When “tyPE” of Cleaning operation setting is set to “nor” (internal timer
only), “HoLd” is displayed
When “tyPE” of Cleaning operation setting is set to “or”, “And”, or “trg”,
“CLn” is displayed
CLn,
HoLd
Setting for
reversing the
logic of hold
command
(cleaning/hold
command)
When the cleaning start is set to “nor”, the logic of the cleaning/hold
command is reversed.
no: Contact is closed when a cleaning command is issued
yES: Contact is opened when a hold command is issued
yES, no
RS-485 setting
RS-485
address setting
Specify the RS-485 communication address.
0 to 99
Display
Item
Description
Setting
range