Measurement
38
Setup menu
The setup menu allows you to set any parameters, select turbidity standard substances for the
detector and units of measurement, and to specify any display, transmission output, relays
and communication addresses.
Entering the setup menu
1.
In the meas mode, hold down HOLD until the HOLD indicator turns ON.
2.
Press
/
to display “SEt” on the main display.
3.
Press ENT.
4.
Press
/
to display a desired item on the sub display.
5.
Press ENT.
Tip
Once the system has entered the hold menu, a measured value changes while a transmission
signal is maintained in a hold state according to the specified method.
Press down the HOLD or ESC to return to the meas mode.
Note
Although the displayed value changes when the HOLD indicator is lit, the transmission output will be
held at the value specified in the setup menu.
Determines
the value
Setup menu
Display of
various settings
Hold mode
Meas mode
Cancels the
value
Changes
the value
Hold down HOLD
or
Hold down ESC