Measurement
37
Calibration data check screen
Calibration data check screen
-
Screen for checking
the light intensity for
zero calibration
Displays the ratio (
) of the factory-
configured amount of transmitted light
during zero calibration to the amount of
transmitted light during the latest zero
calibration.
Note:
This notifies you of any deterioration in the
LED light source.
-
-
Screen displaying
the number of days
elapsed since the
latest zero
calibration
Indicates the number of elapsed days after
the last zero calibration.
Note:
The time counts up every two hours while
the system is energized. The time is not
measured while the power is off.
-
When “yES” is
set in “CAL”
Automatic
calibration wait time
screen
Displays the time until the next calibration is
started in a format of hours-minutes-
seconds.
-
Moving average count check setup screen (damping factor)
-
Moving average
count check setup
screen (damping
factor)
The moving average count can be checked
and changed.
Note:
Increase dF to reduce variations in the
measured values; decrease dF to speed up
the response.
1 to 120
(times)
Displays
Display
condition
Item
Description
Setting range