Installation
22
Range output
For transmission output 2, a maximum of four measurement ranges can be switched
externally or automatically. The range currently output for transmission output 2 can be
recognized at the range output contact.
Tip
An arbitrary range, from range 1 to range 4, can be set for transmission output 2 (zero and span
values). However, it is recommended that you set the lowest value for range 1 and a larger span value
for the rest of the ranges sequentially.
Although a maximum of four transmission ranges can be set, if only two ranges are set
(switched from one to the other), only RNG1 is monitored.
If a contact is to be switched using the external contact, input signals from EXT3 and EXT4.
RNG1
RNG2
Transmission
output 2.A range
RNG1
RNG2
Cur2.A
0
0
Cur2.b
1
0
Cur2.c
0
1
Cur2.d
1
1
1: closed 0: opened