Installation
18
Wire connection
Opening the cover for the converter
1.
Remove the two screw caps located on the top on the front of the converter.
Carefully store the removed caps.
2.
Open the front cover.
3.
Loosen the four screws located on the front of the converter.
Check that these screws are pushed forward by the spring force without falling off.
4.
Open the front case for the converter.
Connection terminal diagram
Note
These screws for the front case and terminal blocks will not drop off by mechanical stopper.
Turn the screws until they are pushed up by spring force for connecting round terminal.
If water droplets are on the case wipe them before opening the case.
Tip
Raise up or remove the black plastic terminal cover hinged on the terminal block for wiring.
RS-485 terminal
Contact input
terminal (M3.5)
Power terminal
(M4)
Detector terminal
(M3)
Contact
output
terminal (M4)
Electric ball valve control output terminal (M4)
Transmission output
terminal (M3.5)
Abnormality alarm output terminal (M4)
Range contact output
terminal (M4)
Solenoid valve control output terminal (M4)