Measurement
43
Calibration menu
Entering the calibration menu allows you to calibrate the residual chlorine sensor and the
temperature sensor.
Entering the calibration menu
There are two different procedures available for entering the calibration mode.
This section describes the procedure for entering calibration menu from the hold mode.
Operation
1.
Hold down HOLD in the measurement mode until the HOLD indicator turns ON.
2.
Use the arrow keys [
/
] and select “CAL” for the calibration menu.
3.
Press ENT.
4.
Use the arrow keys [
/
] to select the desired item displayed on the sub display.
5.
Press ENT.
Tip
Hold down CAL in the measurement mode. This allows you to go directly to the span calibration
mode.
Calibration menu
Display of various
settings
Setting
change
Establishes
the value
Cancels
the value
Press HOLD or ESC
for a while.
Measurement mode
Hold mode
Press HOLD for a while.