Installation
14
Installing the measurement cell
Wall mount or pole mount whichever is available, the pole mount needs another attachment
optionally available.
Note
Install the measurement cell vertically.
The beads cleaning is affected by leaning of the cell, the normal condition for the measurement can
not be maintained.
For installation on the wall
To install the measurement cell on the vertical wall, use the holes in the base shown in the
following figure and also M8 bolts.
RA-10 (overflow type)
RA-20 (inline type)
[Unit: mm]
Front
Rear
Upper maintenance
space: 700 or more
Lower maintenance
space: 1000 or more
2-
10
70
[Unit: mm]
Upper maintenance
space: 500 or more
Lower maintenance
space: 1000 or more
Front
Rear
2-
10
70