4 Various Functions
41
Changing recommended limit times
The recommended limit times of parts are set to the times recommended by HORIBA in
advance.
They can be changed to match the operating conditions with the following procedure.
1.
Press the button of the item to change in the PARTS MANAGEMENT screen.
2.
Press the button on the right of [LIMIT TIME] in the EACH PART MANAGEMENT
screen.
Fig. 44 EACH PART MANAGEMENT screen
Table 20 EACH PART MANAGEMENT screen display items
Tip
The passed time is reset at the time of shipment.
The passed time is not counted during the period from system shutdown to the next startup.
3.
Enter the value on the recommended operation limit setting screen and then press
the [Enter] button.
The setting is updated and the EACH PART MANAGEMENT screen reappears.
The settable range is 0 to 9999 days.
Fig. 45 LIMIT TIME setting screen
Item
Description
[PASSED TIME]
Displays the number of days that have passed since the passed time was reset.
[LIMIT TIME]
Displays the set recommended limit time. When the limit time elapses, an event
occurs.