F-3030 Temperature Bath Addendum: Switching between Internal and External Sensors J810045 (28 Jan 2014)
Note:
The radio button alone in
FluorEssence™ software does not
control switching between internal
and external probes. The temperature
bath’s controller MUST be changed
first.
F-3030 Temperature Bath
Addendum: Switching between
Internal and External Sensors
To switch between internal and external sensors, use the Temperature
Bath’s front panel:
1
Click the right-
arrow ► button
to move to the sub-menu:
2
Scroll up and down with the up
arrow ▲ and down arrow ▼
buttons to choose Settings,
then click the central Enter
button
.
3
Scroll up and down with the up
arrow ▲ and down arrow ▼
buttons to choose Application
Settings, then click the central
Enter button
.