7 DAILY CHECKS
80
7.4
List of Consumables and Replacement Parts
Note
The above replacement period is given as recommended and does not assure any performance.
The replacement period of consumables may become shorter depending upon the installation
environment and operating conditions.
In order to maintain the accuracy, it is recommended that periodical maintenance and checks be
performed when consumables are replaced. For information on maintenance and checks, etc.,
contact us.
No.
Name
Specification
Qty.
Part No.
Replacement
period
1
Filter element
PA-10L
54 mm in diameter
0.5 (t) mm
24 pieces per package
1
3200043947
2 weeks
2
O-ring
JISB2401 G70 (FKM Teflon-coated)
1
3200043785
1 year
3
Filter packing
FKM
1
3200044273
1 year
4
Diaphragm assembly EPDM for GS and GD series, vacuum
2
3200602390
1 year
5
DO unit
For deozonizer
1
3200043513
1 year
6
UV lamp unit
(ozonizer unit)
1
3014038054
1 year
7
UV liner
263
145
0.8 (t) mm, PTFE
1
3200044415
1 year
8
Dehumidifier unit
For APNA-370
1
3200044461
1 year
9
Catalyst tube
For NOx converter
1
3014056464
1 year
10
Air filter
0.3
m
1
3200044257
1 year
11
Scrubber
BAA-050 (activated alumina)
1
3014061530
1 year
12
Silica gel
Containing 500 g
1
3200044017
1 year
13
Pump unit
GD-6EH-10A
1
3200597519
2 years
GD-6EH-23A
1
3200597543
2 years
14
Solenoid valve
WTA-2K-MFF-1
3
3014035610
2 years
15
LCD unit
For APXX
1
3200618696
3 years
16
Battery
CR2032
1
3200043671
3 years