6 FUNCTIONALITIES
50
6.3.2
Analog input
Press the [ANALOG INPUT] button on the MENU/MAINTENANCE screen. The ANALOG
INPUT screen will be displayed.
This screen, which shows the analog input values, is used to check the statuses of analog
signals inputted from the sensor and others.
Fig. 60 ANALOG INPUT screen
The analog input values are listed across 2 pages.
The keys allow you to perform the following operations.
The displayed items and its units are shown below:
[CLOSE]:
Returns to the MENU/MAINTENANCE screen.
[
]:
Displays the previous page.
[
]:
Displays the next page.
Signal name
Unit
Description
SIGNAL(NO)
mV
Voltage of the measured NO value
SIGNAL(NOx)
mV
Voltage of the measured NOx value
DETECTOR
C
Detector temperature
Standard value: 43°C ±5°C
kPa Detector pressure
AMBIENT
kPa Current atmospheric pressure
SAMPLE
L/min
Sample flow rate (optional)
Standard value: 1.1 L/min ±0.3 L/min
DC 24V
V
Power supply voltage inside the APNA-370
Standard value: 24 V ±0.5 V
DC 5V
V
Power supply voltage inside the APNA-370
Standard value: 5 V ±0.5 V