6 FUNCTIONALITIES
65
6.7.1
Machine ID setting
When the [MACHINE ID SETTING] button is pressed on the COMMUNICATION menu, the
MACHINE ID SETTING screen will be displayed.
This screen allows you to set the machine ID that is the machine identification code when RS-
232C and TCP/IP are used.
Note
Set the machine ID in a range between 0 and 249.
The machine ID and the corresponding intended use are shown below.
Fig. 83 MACHINE ID SETTING screen
Immediately after this screen has been displayed, the value for the currently set machine ID is
displayed on the button.
Pressing the MACHINE ID button allows you to change that value. (See "
machine ID" (page 66).)
The keys allow you to perform the following operations.
Machine ID
Intended use
0 to 249
Machine ID that can be set for each equipment piece
255
Machine ID that is used for maintenance and adjustments
* Do not set the machine ID to 255.
[CLOSE]:
Returns to the COMMUNICATION menu without changing the set value.
[SET]:
Saves the changed value. To reflect the changed value, restart is required.