Quality
assurance
User Manual - RAB237AEN
5–27
4.4. Calibration passed
◆
For the instrument to pass calibration, the data must meet the statistical criteria which has 2
conditions:
1-
The coefficient of variations must be within their limits (See
Tab.5-2 Coefficient of variation
2-
The difference between the target value and the average for each parameter calibrated, must be
less than 20%.
Tab.5-2 Coefficient of variation limits
◆
Passed calibration is as follows:
4.5. Calibration failed
◆
For the instrument to fail calibration, the data must meet the statistical criteria which has 2
conditions:
1-
The coefficient of variations are out of their specified limits (See
Tab.5-2 Coefficient of variation
2-
The difference between the target value and the average for each parameter that failed
calibration, is greater than 20%.
◆
Invalid coefficients are displayed in red.
◆
Failed calibration is as follows:
Parameters
WBC
RBC
HGB
HCT
PLT
MPV
CV%
< 2.5
< 2.0
< 1.5
< 2.0
< 5.0
< 3.0
Diag.5-30 Passed calibration
Diag.5-31 Failed calibration
Содержание ABX MICROS ES60
Страница 1: ...Explore the future ABX Micros ES60 HAN706A 2007 HORIBA ABX User manual P n RAB237AEN...
Страница 2: ......
Страница 6: ...User Manual RAB237AEN...
Страница 68: ...ABX Micros ES60 OT CT 3 12 User Manual RAB237AEN...
Страница 116: ...ABX Micros ES60 OT CT 5 18 User Manual RAB237AEN Diag 5 21 LJ graphs printout example...
Страница 128: ...ABX Micros ES60 OT CT 5 30 User Manual RAB237AEN...
Страница 148: ...ABX Micros ES60 OT CT 6 20 User Manual RAB237AEN...
Страница 196: ...ABX Micros ES60 OT CT 7 48 User Manual RAB237AEN...