mmonia electrodes
The structure of the ammonia electrode is shown in Fig. 4.
The ammonia electrode is configured by a glass electrode for pH
measurement, a support pipe, and a membrane cap. The tip of the glass
electrode for pH measurement is made of a pH glass ion selective membrane,
and the tip of the support pipe has a membrane cap provided with a membrane
attached to it. The internal solution is filled between the support pipe and the
glass electrode for pH measurement.
The membrane attached to the membrane cap allows only gases (including
gases dissolved in aqueous solutions) to permeate through.
Fig. 4 Structure of ammonia electrode
Membrane
cap
Membrane
NH
NH
3
(Gas)
(Gas)
NH
3
(Gas)
O-ring
Sample solution
pH glass
ion selective
membrane
Support pipe
Internal solution
Glass electrode
for pH measurement