background image

CODE: GZ0000460289

March, 2017

LAQUAtwin series (Ca

2+

) Operation

1

   

  2017 HORIBA Advanced Techno Co., Ltd.

Instruction Manual (Operation)

COMPACT WATER QUALITY METER
LAQUAtwin-Ca-11

Specifications

Items in package

Consumable parts sold separately

Part Names

Note

Press the switches 0.5 seconds or more unless
otherwise specified.

Initial Setup

Attaching/detaching the sensor

Note

Be careful not to twist the waterproof gasket.

Inserting/removing batteries

Electrode conditioning

Note

Before using the sensor for the first time or
after several days of disuse, perform electrode
conditioning.

Perform calibration after electrode condition-
ing.

1. Place some drops of 2000 ppm standard

solution to the flat sensor.

2. Wait a few hours before use.

There is no need to switch the meter ON.

3. Clean the flat sensor with running water.

Model

LAQUAtwin-Ca-11

Target

Calcium ion (Ca

2+

)

Measurement principle

ISE method

Minimum sample volume

More than 0.3 mL

*1

Measure-
ment range

4 to 9900 [unit: ppm or mg/L]

Resolution 
(default)

Display range: Resolution

4 to 99:

1

100 to 990:

10

1000 to 9900: 100 

[unit: ppm or mg/L]

Calibration

Up to 2 points 
Default: 150 ppm and 2000 ppm

Accuracy

*2

±20% of reading value

Waterproof

IP67 (no failure when immersed in 
water at a depth of 1 m for 30 min) 

*3

Display

Custom (monochrome) digital LCD 
with backlight

Operating 
environment

Temperature: 5

C to 40

C

Humidity: 85% relative humidity or 
less (no condensation)

Power

CR2032 batteries (

2)

Battery life

Approx. 400 h continuous operation 
(backlight off mode)

*4

Material

ABS epoxy (main material)

Dimensions 164 

 29 

 20 mm (excluding projec-

tions)

Mass

Approx. 50 g (excluding batteries)

*1 0.05 mL or more if sampling sheet B (sold sepa-

rately) is used.

*2 The closeness of agreement between a mea-

sured value and an actual value of standard solu-
tion after two-point calibration using 150 ppm and 
2000 ppm standard solutions.

The standard solution used for the later cali-
bration was measured after two-point calibra-
tion at 150 ppm and 2000 ppm.

The calibration and measurement are per-
formed at the same temperature.

The error of standard solutions and rounding
error (

1 digit) are not included.

*3 The meter cannot be used underwater.
*4 When the backlight is used, battery life will

shorten.

Items

Quantity

Sensor

S050

1

Meter

1

Storage case

1

Batteries

CR2032

2

Standard solution

150 ppm

1

2000 ppm

1

Pipette

1

Sampling sheet B (5 sheet-pack)

1

Instruction manual (Operation)

1

Instruction manual (Before use)

1

Items

Specifications

Part No.

Sensor

S050, Ca

2+

3200459869

Standard 
solution

Y051L, Ca

2+

 150 ppm

3200457728

Y051H, Ca

2+

 2000 ppm 3200457727

Sampling 
sheet B

Y046, 100 sheet-pack

3200053858

1

Flat sensor

2

Light shield cover

3

Liquid junction

4

Response mem-
brane

5

Lithium batteries

6

MEAS switch

7

ON/OFF switch

8

CAL switch

9

Waterproof gasket

10 Strap eyelet

Attaching the sensor

1. Power OFF the meter.
2. 
Confirm that the waterproofing gasket is

clean and undamaged.

3. Slide the sensor

onto the meter so
that catch "A" on the
back of the meter fits
into hole "a" on the
sensor tongue as
shown.

1

2

3

4

5

6

7 10

8

9

a

A

Waterproof
gasket

Detaching the sensor

1. Power OFF the meter.
2. 
Lift the sensor tongue

tip and slide the sensor
a little away from the
meter.

3. Pull out the sensor all the way from the

meter.

Inserting the batteries

1. Power OFF the meter.
2. 
Slide both batteries into

the battery case as
shown.

Be sure to use two
CR2032 batteries, and put
them with the plus sides
(+) upwards.

Removing the batteries

1. Power OFF the meter.
2. 
Use a ball-point pen or

other tool to pry the batter-
ies out from the clips as
shown.

Battery

Battery clip