Ion concentration measurement
39
●
Setting the instrument
• Setting the temperature display
Because the ion electrode is not equipped with a temperature sensor, the temperature
electrode must be connected in order to use the automatic temperature measurement
function. When measuring ion concentration, the temperature sensor on the electrode
measures the sample temperature and displays it on the instrument. The sample
temperature is measured by the temperature electrode during the ion concentration
measurement, and the measured value is displayed on the instrument. If automatic
temperature measurement is not used or a temperature connector (T) is not connected
to the instrument with ATC set, arbitrary temperature set in the instrument is displayed.
This cannot be set separately from the automatic temperature compensation setting for
pH and the automatic temperature measurement setting for mV (ORP).
1. Press the
key to enter the setting mode.
2. Press the
keys to select "TEMP" (temperature setting) and then press the
key.
If the temperature connector (T) is connected, "ATC" (automatic temperature
compensation) appears. If the temperature connector (T) is not connected, "MTC"
(manual temperature compensation) appears.
In the case of "ATC", the temperature sensor can be calibrated (refer to "Calibrating
temperature sensor" (page 63)).