Dissolved oxygen measurement
34
• Setting air pressure correction (default: 1013 hPa)
Amount of oxygen that can be dissolved in water is affected by air pressure in the
location where measurement is taking place. You need to set the air pressure of the
measurement location to the instrument before calibration to correct the pressure
influence, and then perform measurement.
1. Press the
key to enter the setting mode.
2.
Press the
keys to select the "PRES" (air pressure collection setting) and
then press the
key.
3. Enter the air pressure value of the measurement location and then press the
key.
4. Press the
key to reflect the change.
key
key
Содержание LAQUAact D-75G
Страница 2: ......
Страница 22: ...M E M O...
Страница 88: ...M E M O...
Страница 94: ...M E M O...
Страница 116: ......
Страница 117: ......
Страница 118: ......