Dissolved Oxygen Calibration & Measurement

1. Rinse the DO electrode 

with clean water and wipe 

with lint-free tissue.

2. Place the DO electrode in water vapor-

saturated air (calibration bottle with moist 

sponge) or in air-saturated water.

3. Press CAL button on the meter to switch to 

calibration mode.

4. Press ENT button to confirm the DO reading. To abort calibration, press 

MEAS button.

Make sure that the DO cap and temperature 

sensor of the DO electrode are fully 

immersed in the sample. 

5. Immerse the DO electrode in zero solution (0.2% sodium sulfite) and repeat 

step nos. 3 - 4 to perform zero calibration (optional).

6. Rinse the DO electrode with clean water and wipe with lint-free tissue.

7. Immerse the DO electrode in sample.

Make sure that the DO cap and temperature sensor of the DO electrode are fully immersed in the zero 

solution. 

In step no. 3, CAL button must be pressed twice so that the meter will switch to zero calibration mode. 

Quick Guide

LAQUA WQ-300 Series 

Handheld Meters

www.horiba-laqua.com

The meter will show DONE 

briefly then switch to 

measurement mode. The 

calibrated DO value and the 

electrode icon will appear in 

measurement mode.

Press

There should be no water droplets on 

the DO cap and temperature sensor.

        P/N: 3200832950 

 

 

 

 

 

 

     CODE: GZ0000584707

Press

Two calibration modes are available in the 

DO meter:
1. 

DO saturation (%)

2. 

DO concentration (mg/L)

The default mode is the former. Press 

MODE button to switch to the latter.

100% is displayed at the bottom of the screen. The 

 icon 

on the screen and the LED light on the electrode (if switched 

on) will blink until the reading is stable.