![Horiba Scientific DS-72G Скачать руководство пользователя страница 62](http://html1.mh-extra.com/html/horiba-scientific/ds-72g/ds-72g_instruction-manual_2162306062.webp)
42
3.2.4 Temperature Conversion Function Setting
HINT!
When select the pure water mode or the natural water mode, the temperature conversion
conforms to the following standards.
Pure water: ASTM D 1125-91 Table3
Natural water: ISO7888:1985 (JIS K0400-13-10:1999)
Input Temperature Conversion Factor
Reference Temperature Setting of Temperature Conversion
1.
Tap
on the right of the TEMP conversion
item.
2.
Select the temperature conversion method.
1.
Tap
on the right of the TEMP conversion
item.
2.
Select "Manual" on the TEMP conversion
screen.
3.
Tap
on the right of the Coefficient item.
4.
Enter the temperature conversion factor on the
numerical-key screen.
5.
Tap .
The setting applies.
To cancel the settings, tap
.
1.
Tap
on the right of the TEMP conversion
item.
2.
Select "Manual" on the TEMP conversion
screen.
3.
Tap
on the right of the Reference item.
4.
Enter any temperature value between 15°C and
30°C on the numerical-key screen.
5.
Tap .
The setting applies.
To cancel the settings, tap
.
The measured COND value of a sample varies
with the temperature. In addition, the change
degree with temperature depends on the sample
property.
If the change degree (temperature coefficient) of
the sample is known, set this item to ON to
display COND values converted at 25
C.
If the temperature coefficient is unknown, set this
item to OFF.
Содержание DS-72G
Страница 1: ...Instruction Manual COND METER DS 72G...
Страница 2: ......
Страница 12: ...For Your Safety X JPN EN61326 1 EN61010 1 II 2...
Страница 20: ......
Страница 104: ......
Страница 105: ......
Страница 106: ......