Wipe off the moisture.
Wipe off the moisture using soft and clean lint-free paper or cloth.
7
Open the rubber stopper of the
internal solution fill port.
If the ion selective membrane is rubbed severely to remove
the moisture, the ion selective membrane may tear or static
electricity may be generated, resulting in a measurement
errors. Press the cloth against the ion selective membrane
lightly so as to soak up the water.
Conditioning of the electrode
In the following cases, perform
conditioning of the electrode before
calibration.
Before the first use
Before use after storage for two weeks or
longer
If the electrode is less responsive
An electrode is conditioned by soaking it in a standard solution
for 1 hour or longer.
Use a standard solution whose concentration is 100 mg/L or
more.
A standard solution with TISAB added can not be used.
Note
Note
Rubber stopper
Note
Note
Standard
solution