15
Electrode Preparation
Calibration
Measurement
After Measurement
R
insing the electrode
Make sure that no dirt and other foreign matters have
adhered to the electrode before measurement.
If dirt and other foreign matters have adhered to the
electrode, rinse the electrode as described in the following
procedure.
1
Rinse the electrode with pure water.
2
Wipe off the moisture.
Wipe off the moisture using soft and clean lint-free paper or cloth.
If the ion selective membrane is rubbed severely to remove
the moisture, the ion selective membrane may tear or static
electricity may be generated, resulting in a measurement
errors. Press the cloth against the ion selective membrane
lightly so as to soak up the water.
View video
View video
Note
Note