5.3.11
VA - Pairs of values
Menu -> Settings -> Sensor -> VA Input -> Pairs of values
If pairs of values is selected (see 5.3.7.1), then the values stored here for determining the measured
value are applied as a calibration curve.
Selection:
Configuration
= enter the
quantity
of pairs of values (minimal 2, maximal 30) for
processing and display
Pairs of values
= display and change of pairs of values.
A pair of values always consists of a velocity value in m/s
and a frequency value in Hz.
The condition being: the pairs of values must be ever
increasing, i.e. the next velocity and frequency value must always
be greater than the previous one.
Example for quantity = 03
01:000.50m/s, 00010Hz
02:010.00m/s, 00350Hz
03:040.00m/s, 01770Hz
If the frequency measured value is greater than in the last pair of values, then the velocity value is
calculated. However, this means that the measurement uncertainty increases as this value is then outside
the calibrated range.
Settings for thermal sensors TA:
5.3.12
TA - Basic settings
Menu -> Settings -> Sensor -> TA Input ->
Basic settings
The connected
thermal sensor
determines the parameter type.
5.3.12.1
Type
Setting the type of thermal sensor:
Refer to the relevant technical documents for selection options.
Selection:
KKZ
= the
calibration number (KKZ)
is individually determined for each
sensor and modifies the basic characteristics.
KKZ:
enter as a 14-digit figure, in which each digit has a
0..9..A..F range (hexadecimal = 16 possible variables).
The actual KKZ can be found in the technical data sheet, calibration
certificate or directly on the sensor
pairs of values
= special calibration characteristic specifically matched to the
measuring task based on up to 30 supporting points. Entering or
amending the points, see under 5.3.15
The actual pairs of values can be found in the technical documents.
5.3.13
TA - Measurement settings
Menu -> Settings -> Sensor -> TA Input -> Measurement settings
The task to be carried out determines the parameter, section and profile factor to be set.
5.3.13.1
Section
Setting the measuring section for measuring in pipelines for flow rate display:
Selection:
Circular
= for pipes with circular section
di/mm
: enter the ID in mm
Rectangular
= for pipes with rectangular section
a/mm
: enter the inner surface a in mm
b/mm
: enter the inner surface b in mm