Analog input settings:
5.3.17
Analog input
Menu -> Settings -> Sensor -> Analog input
Input:
Selection:
4-20mA = current input 4-20 mA is selected
0-10V = voltage input 0-10 V is selected
Designation:
for each of the two inputs a designation of up to 13 digits for the sensor can
be entered.
Example: pressure sensor
Equivalency:
for each of the two inputs a display equivalency can be defined. For this
purpose the desired initial value of 4 mA or 0 V and the desired final value of
20 mA or 10 V is entered respectively.
Example: 4 ... 20 mA -> 900 ... 1600 hPA
Unit:
for each of the two inputs a unit of up to 5 digits for measured value display
can be entered.
Example: hPa
(unused digits are marked with an "*" and are not shown in the display
and data logger)
Counter settings:
5.3.18
Counter
Menu -> Settings -> Counter
Saettings for the counter,
5.3.18.1
Reset
Resets counter to 0
5.3.18.2
Decimals
Determines the number of decimal places (0, 1 or 2) for display
Display settings:
5.3.19
Display
Menu -> Settings -> Display
Settings for measured value display. Determines the quantity of the simultaneously displayed values /
input channel and assignment of these values to the 3 channels.
5.3.19.1
Sensor
Selection:
Sensor:
= flow sensor selection
FA: = vane wheel sensor
VA: = vortex sensor
TA: = thermal sensor
Only the selected sensor may be connected!
5.3.19.2
Rows
The number of rows determines in how many input channels (1, 2 or 3) the measured values are
displayed simultaneously after switch on.