Page 44
ST 700 Series HART/DE Option User’s Manual
Revision 4.0
Online/Device
Setup/Detailed Setup
(
Continued …)
Output Condition
The following output-related values and operations are
available in this menu:
Poll Address:
Click HART short address 0 to 63.
Num req preams:
Displays the number of
required request preambles for the ST 700 HART
communications
PV AO:
Displays the current value of AO
(analog output) in milliamps
PV AO Alm Type:
Displays the current position
of the failsafe jumper on the electronics board
(upscale/ lowscale burnout)
Loop mA:
Click the Loop Current Mode
configuration:
Enable:
Enables loop current mode (analog
output will operate as a 4 to 20 mA signal
consistent with the transmitter output)
Disable:
Disables loop current mode (analog
output will be fixed at 4 mA)
Loop test:
This function enables the user to test
the Analog Output measurement at any value
over the full operational range. Click a current
value to apply and verify the current output on
the loop with a calibrated meter.
This function is only available when
Loop mA
(Loop Current mode) is Enabled.
D/A trim:
perform an analog output calibration at
4.00 and 20.00mA (0% and 100% output).
Online/Device
Setup/Meterbody
Selections
Displays detailed information pertaining to the
Meterbody, including construction configuration,
materials of construction, fill fluids, and process
connection details.
Online/Device
Setup/Display Setup
(for Basic Display &
Standard Display
Only*)
Display
Connected
Indicates the type of display installed on the transmitter.
Read Screen Info
Displays configured information for each of the eight
display screens. Click a screen number to view details
of the configurations selected in the
Screen
Configuration
menu.
Screen
Configuration
The display screen may be customized with the following
settings:
PV Selection:
Click the displayed process
variable as:
o
Differential Pressure
o
Loop Output
o
Percent Output
Pressure Units:
Click desired standard
pressure unit.
Number of Decimals:
Click the number of
decimal places for the PV display from none to 3.
Custom Tag:
Enter a custom tag name for the
screen title up to 14 alphanumeric characters.
Содержание SmartLine ST700
Страница 88: ...Page 74 ST 700 Series HART DE Option User s Manual Revision 4 0 This page has been intentionally left blank ...
Страница 98: ...Page 84 ST 700 Series HART DE Option User s Manual Revision 4 0 This page intentionally left blank ...
Страница 115: ...Revision 4 0 ST 700 Series HART DE Option User s Manual Page 101 11 9 Trend Charts ...
Страница 175: ...Revision 4 0 ST 700 Series HART DE Option User s Manual Page 161 This page intentionally left blank ...
Страница 188: ...Page 174 ST 700 Series HART DE Option User s Manual Revision 4 0 20 9 Trend Charts ...