![Honeywell SC500 Скачать руководство пользователя страница 49](http://html.mh-extra.com/html/honeywell/sc500/sc500_manual_151996049.webp)
SC Series Instruction Manual - Model SC500 page 49
Limi
ts
9
9.4.6
LIMIT.LATCHING
Menu Item
This specifies whether to latch the activated limit so that only manually clearing the
limit will deactivate it. A latched limit can be manually cleared by pressing the
[CLEAR] button or sending the “F8” serial communications command.
The options for this menu item are:
•
“
ON
” means that once a signal activates the limit, the limit will remain activated until
it is reset.
•
“
OFF
” means that the limit will deactivate immediately when the signal crosses the
return point.
9.4.7
LIMIT.CHANNEL
Menu Item
This specifies the channel that is monitored by this limit.
9.4.8
LIMIT.SOURCE
Menu Item
This designates the data source of the channel monitored by this limit. Each channel
has three data sources: the live tracking value (TRACK), its highest value (PEAK),
and its lowest value (VALLEY).
The options for this menu item are:
•
“
TRACK
” means the live tracking value of the channel.
•
“
PEAK
” means the highest value of the channel since the peak/valley detector was
last cleared.
•
“
VALLEY
” means the lowest value of the channel since the peak/valley detector was
last cleared.
9.4.9
LEAVE MENU
Menu
Item
When this menu item is displayed, press [ENTER] to leave this menu. Press [EXIT] to
return to the RUN mode.
Содержание SC500
Страница 8: ...page 8 008 0671 00 ...
Страница 14: ...page 14 008 0671 00 ...
Страница 18: ...page 18 008 0671 00 ...
Страница 24: ...page 24 008 0671 00 ...
Страница 38: ...page 38 008 0671 00 ...
Страница 42: ...page 42 008 0671 00 ...
Страница 50: ...page 50 008 0671 00 ...
Страница 72: ...page 72 008 0671 00 ...
Страница 82: ...page 82 008 0671 00 ...
Страница 90: ...page 90 008 0671 00 ...
Страница 116: ...page 116 008 0671 00 ...
Страница 126: ...page 126 008 0671 00 ...