Honeywell SC1000 Скачать руководство пользователя страница 1

Models SC1000 & SC2000

Model SC3004

SC Instrumentation

Signal Conditioning,
Self-Calibrating
Digital Indicators

Sensotec Sensors

Model SC2001

Sensing and Control

Содержание SC1000

Страница 1: ...Models SC1000 SC2000 Model SC3004 SC Instrumentation Signal Conditioning Self Calibrating Digital Indicators Sensotec Sensors Model SC2001 Sensing and Control ...

Страница 2: ... is advised that if the equipment is used in a man ner not specified in this manual the protection provided by the equipment may be impaired CAUTION Only qualified service trained personnel who are aware of the hazards involved should remove the cover from the instrument or connect external wiring to the instrument WARNING PERSONAL INJURY DO NOT USE these products as safety or emergency stop devic...

Страница 3: ...he Correct Channel of the Instru ment 15 2 3 Turn on the Instrument 16 2 4 Use the SETUP Menus to Enter Transducer Information 17 2 5 Calibrate the Transducers to Their Channels 17 2 6 The SC Series Instrument is Ready for Use 17 Chapter 3 Operating Modes 19 3 1 Operating Modes 19 3 2 INITIALIZE Mode 19 3 3 RUN Mode 20 3 3 1 Display 20 3 3 2 VALUE button 20 3 3 3 CLEAR Button 20 3 3 4 CHANNEL butt...

Страница 4: ...on 32 4 4 8 Fuse Replacement 32 4 5 Model SC3004 33 4 5 1 External Arrangement 33 4 5 2 Rear Panel 33 4 5 3 Panel Mounting 33 4 5 4 Bench Mounting 33 4 5 5 Case Removal 34 4 5 6 Rear Panel 34 4 5 7 Internal Arrangement 34 4 5 8 Cleaning 34 4 5 9 Fuse Replacement 34 Chapter 5 System Connector 35 5 1 Introduction 35 5 2 System Connector Pinout 36 5 3 Function Input Pins 37 5 3 1 Overview 37 5 3 2 Ex...

Страница 5: ...51 8 2 3 LOWER MODE Menu Item 51 8 2 4 DISPLAY DISABLE Menu Item 51 8 3 Menu Items for Model SC3004 51 8 3 1 POWER ON CHANNEL Menu Item 51 Chapter 9 Limits 53 9 1 Understanding Limits Set Points and Return Points 53 9 2 Limit Operation 54 9 2 1 Actions When Activated 54 9 2 2 Scan Time 54 9 3 Limit Menus 55 9 3 1 Overview 55 9 3 2 LIMIT ENABLE Menu Item 55 9 3 3 LIMIT SETPOINT Menu Item 55 9 3 4 L...

Страница 6: ...FUNCTION Menu Items 85 11 5 4 CALIBRATION TYPE Menu Item 87 11 5 5 CALIBRATION DATA Sub Menu 88 11 5 6 CALIBRATE Menu Item 89 11 5 7 DAC SETUP Sub Menu 91 11 5 8 DIAGNOSTICS Sub Menu 92 11 6 Electrical Null and Transducer Mounting 93 11 6 1 Overview 93 11 6 2 Procedure 93 11 7 Analog Output Configuration 94 11 7 1 Identifying the Output Type 94 11 7 2 Channel Menu Items 94 11 7 3 Output Selection ...

Страница 7: ...eshooting 122 12 8 1 Error Messages 122 12 8 2 Common Problems and Solutions 122 Chapter 13 Relay Output Channel 123 13 1 Features 123 13 1 1 First Channel Installed 123 13 1 2 Second Channel Installed 123 13 1 3 Third Channel Installed 123 13 1 4 Fourth Channel Installed 123 13 2 Wiring 124 13 3 Setup Procedure 125 13 4 Specifications 126 13 5 Channel Menu 127 13 5 1 DIAGNOSTICS Sub Menu 127 Chap...

Страница 8: ... 16 3 Setup Procedure 145 16 4 Channel Menu 146 16 4 1 SENSOCODE P N Menu Item 146 16 4 2 USER VALUES Sub Menu 146 16 4 3 VIEW SENSOCODE Menu Item 146 16 4 4 VIEW COMMANDS Sub Menu 146 16 4 5 DISPLAY SETUP Sub Menu 147 16 4 6 POWER ON SOURCE Menu Item 148 16 5 Troubleshooting 149 Chapter 17 Error Messages 151 17 1 Overview 151 17 2 Error Message List 152 Chapter 18 Setup Menu Reference 157 18 1 Na...

Страница 9: ...plains the differences between the SC1000 SC2000 SC2001 and SC3004 chassis Information relating to the hardware chassis such as panel and rack mounting is given Chapter 5 System Connector contains information about wiring to the 25 pin System connector to access the Limit Outputs Function Inputs and serial commu nications Chapter 6 System Menu discusses the System Menu which allows you to examine ...

Страница 10: ...le additional limits in the instrument Chapter 14 DAC Output Channel explains the configuration and operation of additional digital to analog voltage or current outputs Chapter 15 Split Display Virtual Channel shows how you can display two values from any of the channels in the instrument at the same time Chapter 16 Mathematics Virtual Channel describes the flexibility of customer specific SensoCo...

Страница 11: ...ations Guide document 008 0610 00 describes in detail how to communicate with an SC Series instrument using RS 232 and RS 485 Wiring diagrams sample programs and descriptions of each command are included A printed copy of this document is available for order or you may download it from http www honeywell com sensotec Supplemental Instructions If an instrument is configured with Mathematics Virtual...

Страница 12: ...ilable SC1000 1 to 4 physical channels 3 8 DIN case no limits or peak detector SC2000 1 to 4 physical channels 3 8 DIN case SC2001 1 to 4 physical channels portable case SC3004 1 to 14 physical channels 19 rack mount case 1 to 3 quad line displays 1 3 3 Channel Types Channels can be one of three types Input Output or Virtual Input Channels Input channels are hardware circuit boards with a unique c...

Страница 13: ...troduction 1 Mathematics Virtual channels run small programs written in an interpretive language called SensoCode This provides great flexibility which allows the SC Series to do many jobs which otherwise requires a personal computer or PLC ...

Страница 14: ...quire ments A Factory Calibration Sheet Data area unalterable by the customer can be cop ied back into the User Calibration Data 1 4 3 Information Stored The following information is stored inside transducers equipped with Signature Cali bration Full scale mV V The full scale millivolt per volt mV V rating of the transducer when its full load is applied also called calibration factor Shunt Cal mV ...

Страница 15: ...e of Calibration The Customer Information Sheet that shipped with your instrument This sheet describes which cards are installed in each channel Power cord for the instrument 2 2 Connect the Transducer to the Correct Channel of the Instrument For each transducer attach its connecting cable to the transducer and then to the correct 12 pin channel connector on the instrument The Customer Information...

Страница 16: ...ue lowest value seen since the CLEAR button was last pressed Data value 00000 displays the value from the transducer in engineering units Units label PSIG indicates the engineering units being used up to 4 charac ters Tare indicator A symbol indicates tare is off a Ë symbol indicates tare is on NOTICE If the channel s display flashes APPLY 00000 or some other load value the instrument has detected...

Страница 17: ...to Their Channels You can skip this step if You are using a transducer equipped with Signature Calibration or The transducer and instrument were purchased together and set up by Sen sotec Otherwise you must choose a calibration method e g Shunt Calibration mV V Calibration or Known Load Calibration appropriate for your application and use the SETUP menu for that channel to calibrate the channel to...

Страница 18: ...page 18 008 0608 00 ...

Страница 19: ...ode When the instrument is powered up or otherwise reset it enters the INITIALIZE mode As the instrument enters this mode all segments of the display and all front panel indicator lights if available illuminate momentarily Next each channel in the instrument is checked for proper operation If a problem is detected the instrument may enter the ERROR mode Depending on the type of channel other actio...

Страница 20: ...000 SC2000 SC2001 The contents of the display s lower line is selected with the DISPLAY MENU LOWER MODE menu item as either displaying a channel or indicating the status of Limits See Indi cator Lights on page 21 3 3 2 VALUE button After the channel number the next characters indicate which data value for the dis played channel is shown There are three possibly four data values available from each...

Страница 21: ...will illuminate when tare is on To tare all channels simultaneously use the System connector s Function 1 Input pin See System Connector on page 35 To tare a single channel regardless of which channel is being monitored by the display use the channel s AUX1 or AUX2 control pins on its connector See the particular chapter regarding that channel for details 3 3 6 Indicator Lights ModelsSC1000 SC2000...

Страница 22: ... short description of the error The first two digits xx hold the error code The last two digits yy is channel number that caused the error For example ERROR 60 ON CH 01 indicates that error number 60 occurred on channel 1 While the instrument is in the ERROR mode no other operations are taking place except for limited serial communications capabilities See Error Message List on page 152 for a list...

Страница 23: ...SETUPMenusandMenu Items The following table shows how to move though menus and menu items and how to display or change the setting of a particular item To change the setting of a menu item Press ENTER to display the current setting which is preceded by the character Press UP or DOWN until you see the setting you want To abandon changes press EXIT To accept the new setting and move to the next item...

Страница 24: ...page 24 008 0608 00 ...

Страница 25: ...rtual channels Input channels and Output channels are printed circuit boards that occupy a phys ical slot inside the instrument s chassis 12 pin connectors are located on the rear panel to connect to each Input or Output Channel Each channel in the instrument is identified by a number channel 01 channel 02 etc A Virtual channel exists in software only it does not occupy a physical slot inside the ...

Страница 26: ... 30º C to 90º C 30º C to 90º C 30º C to 90º C 30º C to 90º C Operating temp 5º C to 40º C 5º C to 40º C 5º C to 40º C 5º C to 40º C Relative Humidity 80 max for temperatures up to 31º C decreasing linearly to 50 max at 40º C Other indoor use at altitudes up to 2000 m Pollution Degree 2 Overvoltage Cate gory II SPECIAL FEATURES Limits Quantity N A 4 std 16 max 4 std 12 max 4 std 16 max Digital isol...

Страница 27: ... is given below Figure 4 1 External Arrangement of AC powered SC1000 and SC2000 4 3 3 Rear Panel The pinout for the 25 pin System connector is provided later in this chapter The pinouts for the individual channels are located in the chapter for that channel 4 3 4 Panel Mounting The panel cutout size conforms to the 3 8 DIN standard Panel Mounting Jacks order code AA928 are available that slide int...

Страница 28: ...g Jacks into the slide slots Slide them as far toward the panel as possible Step 6 Reinsert the setscrews and tighten them This will force the Panel Mounting Jacks toward the rear side of the panel drawing the instrument tightly into place 4 3 5 Rack Mounting A Rack Mounting Kit order code AA934 is available for mounting a single SC1000 or SC2000 into a 19 2U rack It includes the Panel Mounting Ja...

Страница 29: ...the four black Phillips head machine screws that secure the back panel of the instrument to the case These screws are located on the rear of the case one at each of the four corners Do not remove the cable shield connection screw in the center Step 4 The circuit boards will slide out of the rear of the case as a unit 4 3 8 Internal Arrangement The figure below shows the names and locations of the ...

Страница 30: ... cloth or tissue and a mild cleaner Do not use liquid or aerosol cleaners Do not allow any cleaner inside the instrument 4 3 10 Vehicle Power Option Model SC1000 and SC2000 instruments are available with a vehicle power option for operation with batteries and linear DC power supplies See Specifications on page 26 voltage and power requirements 4 3 11 Fuse Replacement The power line fuses of AC pow...

Страница 31: ...Panel The pinout for the 25 pin System connector is provided later in this chapter The pinouts for the individual channels are located in the chapter for that channel 4 4 4 Case Removal Step 1 Remove the 10 Phillips head machine screws that secure the front panel to the case 0 1 2 3 4 5 6 7 8 9 A B C D E F 0 1 2 3 4 5 6 7 8 9 A B C D E F WARNING Disconnect the power cord and all cables from the in...

Страница 32: ...cleaner inside the instrument 4 4 7 Vehicle Power Option The Model SC2001 is available with a vehicle power option for operation with bat teries and linear DC power supplies See Specifications on page 26 voltage and power requirements 4 4 8 Fuse Replacement The power line fuses of AC powered instruments are located within the instru ment s power entry module on the rear panel Use two 2A 250V fast ...

Страница 33: ...t channel 4 5 3 Panel Mounting The panel space necessary conforms to the EIA 19 rack mount standard Panel mounting ears are attached to the instrument 4 5 4 Bench Mounting Panel mounting ears are attached to the instrument and may be removed if they are not needed 0123456789ABCDEFGHIJ 0123456789ABCDEFGHIJ 0123456789ABCDEFGHIJ 0123456789ABCDEFGHIJ 0123456789ABCDEFGHIJ 0123456789ABCDEFGHIJ 012345678...

Страница 34: ...ser installable printed circuit boards will slide out of the rear of the case once the case and rear panel have been removed as described above 4 5 8 Cleaning Turn off the instrument and unplug all connectors Use a soft cloth or tissue and a mild cleaner Do not use liquid or aerosol cleaners Do not allow any cleaner inside the instrument 4 5 9 Fuse Replacement The power line fuses are located with...

Страница 35: ... D subminiature System connector is used for the following Communication by RS 232 or RS 485 RS 232 DCE standard designations have been maintained Digital Function Inputs such as for resetting tare peak and latched limits Open collector digital Limit Outputs for limits 1 through 4 not available on Model SC1000 ...

Страница 36: ...GND RS 232 RS 485 reference Reference 8 DCD RS 232 Data Carrier Detect not connected N A N A 9 FUNC 2 Function Input 2 Clear Peak Valley Limits Input 19 10 FUNC 1 Function Input 1 Tare Off for all channels Input 19 11 FUNC 0 Function Input 0 Input 19 12 RS 485 TB RS 485 Transmit B Output 7 13 RS 485 TA RS 485 Transmit A Output 7 14 L1 Limit 1 Output Open Collector Output 19 15 L2 Limit 2 Output Op...

Страница 37: ...tion Input pins perform the default actions described in the System Connector Pinout on page 36 However a SensoCode program running on a Mathe matics Virtual Channel may replace these default actions Consult the Customer Information Sheet included with your instrument for details 5 3 2 Example For example assume that you wish to use Function Input 3 to tare all channels in the instrument simultane...

Страница 38: ...running on a Mathematics Virtual Channel may over ride this behavior Consult the Customer Information Sheet included with your instru ment for details 5 4 2 Example For example assume that a remote indicator is to be lighted when Limit 1 is activated An external 24 VDC power supply connects to the indicator When Limit 1 is deactivated there is high resistance between the Limit 1 Output pin and the...

Страница 39: ...SC Series Instruction Manual page 39 System Connector 5 ...

Страница 40: ...page 40 008 0608 00 ...

Страница 41: ...sage NOT INSTALLED is shown Press ENTER again to return to the sub menu listing of all channels 6 2 3 DIAGNOSTICS Sub Menu This sub menu allows exercising and monitoring of the System connector s output and input pins OUTPUT n PIN nn Menu Items These menus items are used to select an output pin to turn on connected to pin 19 or off disconnected from pin 19 The output pins are updated immediately I...

Страница 42: ...will be installed as the next channel number 2 Examine the circuit card to be installed and orient it as shown below On it you will find the address jumper block Figure 6 1 Address Jumper Block Location 3 Change both address jumpers to match the next available channel in the instrument according to the chart below Do not skip any channel numbers Figure 6 2 Address Jumper Settings NOTICE Installing...

Страница 43: ...m mence 12 If the installation was successful DONE will be displayed If the installation failed you will see one of the following messages SYSTEM IS FULL There are no unused channels available in the instrument WON T INSTALL The Output Relay card type you selected to install does not match the next available set of Limits For example the Limit 09 12 card type will not install unless the Limit 05 0...

Страница 44: ... chassis See Case Removal on page 28 for SC1000 and SC2000 instruments See Case Removal on page 34 for SC3004 instruments 6 2 6 DEFAULT CHANNEL Menu Item This menu item will reset all settings for the channel you select to their factory default values Before the channel settings are reset you are asked ARE YOU SURE To cancel this operation press the EXIT button If YES is selected the channel s set...

Страница 45: ...SC Series Instruction Manual page 45 System Menu 6 ...

Страница 46: ...page 46 008 0608 00 ...

Страница 47: ...ruction manual is available to assist with the hookup and wiring for data communications as well as provide detailed information of all the available commands Ask for manual 008 0610 00 SC Series Communications Guide or download it from http www honeywell com sensotec 7 2 Wiring The System connector on the instrument s rear panel is used among other things for serial communications See System Conn...

Страница 48: ...ssages in parallel on the line the so called multi drop system To avoid garbled transmissions only one device should respond to a particular message Therefore every device on the loop must have a unique address If you have an instrument with more than one channel transmission must be addressed to the appropriate channel within the instrument 7 3 2 Parameters All SC instruments use no parity 8 data...

Страница 49: ...ial communications The default baud rate is 9600 7 4 5 AUTO LINE FEED Menu Item This allows you to examine the setting on off of the auto line feed function When set to on a line feed character ASCII code decimal 10 is transmitted just before the carriage return ASCII code decimal 13 that signifies the end of the instrument s response If the instrument is connected to a dumb terminal or printer in...

Страница 50: ...page 50 008 0608 00 ...

Страница 51: ...menu item is set to CHANNEL To change whether the channel s track peak or valley data value is displayed on power up use the channel s OPERATION POWER ON SOURCE menu item 8 2 3 LOWER MODE Menu Item Determines what is displayed on the lower line The choices are BLANK means that nothing is displayed LIMIT 01 04 means that the status of Limit 1 Limit 2 Limit 3 and Limit 4 are displayed as described i...

Страница 52: ...page 52 008 0608 00 ...

Страница 53: ...signal level is changing slowly near these points chatter may result and the limit rapidly ener gizes and de energizes To keep this from happening we can specify the SET POINT at which the action would occur as well as the RETURN POINT where we want the action to cease If the amount of dead band between these two values is more than the expected noise the desired action will take place only once T...

Страница 54: ... will illuminate The relay for the limit on this Relay Output channel will activate A Relay Output channel can be installed for Limits 9 10 11 and 12 When these limits are activated the following actions occur If the Relay Output channel is displayed the display will indicate the status of these limits The relay for the limit on this Relay Output channel will activate A Relay Output channel can be...

Страница 55: ...meric value at which the limit deactivates In general this number should differ from the set point values by at least 1 of full scale Carefully consider what value should be entered for the RETURN POINT If the signal is expected to approach the SET POINT from a lower value the RETURN POINT value should be less algebraically than the SET POINT If the signal is expected to approach the SET POINT fro...

Страница 56: ...ge 56 008 0608 00 Figure 9 1 Limit Operation when LIMIT ENERGIZE SIGNAL SETPOINT Figure 9 2 Limit Operation when LIMIT ENERGIZE SIGNAL SETPOINT Figure 9 3 Limit Operation when LIMIT ENERGIZE SIGNAL INSIDE ...

Страница 57: ...it is reset OFF means that the limit will deactivate immediately when the signal crosses the return point 9 3 7 LIMIT CHANNEL Menu Item This specifies the channel that is monitored by this limit 9 3 8 LIMIT SOURCE Menu Item This designates the data source of the channel monitored by this limit Each channel has three data sources the live tracking value TRACK its highest value PEAK and its lowest v...

Страница 58: ...page 58 008 0608 00 ...

Страница 59: ...t resolution at the slowest filter setting across a wide variety of mV V input ranges Three methods of calibrating the Strain Gage Input channel to the transducer are available known load calibration shunt calibration and mV V calibration The benefits of each are discussed in CALIBRATION TYPE Menu Item on page 67 Two rear panel control inputs can be field configured for such functions as remote ta...

Страница 60: ...Figure 10 1 Unamplified Transducer Connection to Strain Gage Input Channel Table 4 3 Strain Gage Input Channel Pin Connections Pin Label Function Reference Pin 1 top EXC Excitation 10 2 SEN Sense 10 3 SEN Sense 10 4 EXC Excitation 10 5 SIG Signal 10 6 SIG Signal 10 7 OUT Analog Output 8 8 OUT Analog Return 9 MEM Signature 10 10 MEM Signature Digital Ground 11 AUX1 Auxiliary Function 1 connect to p...

Страница 61: ...cribed in CALIBRATION TYPE Menu Item on page 67 It is important to know your application in order to select the appropriate calibration type Step 3 Enter the CALIBRATION DATA If your transducer has Signature Calibration you don t need to enter these values as they are entered automatically If your transducer was ordered and shipped along with the instrument you don t need to enter these values as ...

Страница 62: ...Response see FREQ RESPONSE Menu Item on page 63 INSTRUMENT ONLY ACCURACY Sense wires used Frequency Response setting 16Hz Linearity repeatability hysteresis mV V Calibration 0 1 F S Known Load Calibration 0 01 F S AUXILLIARY INPUTS Quantity 2 Type momentary contact closure Response Time 5ms Field Selectable Functions tare on tare off peak valley clear peak valley hold track hold ANALOG OUTPUT Outp...

Страница 63: ...re negative than 10 display counts for any 10 second period For example assume that the DISPLAY DECPT menu item is set to 000 00 two deci mal points If the tracking value never leaves the range of 000 10 to 000 10 for any 10 second period then tracking value will reset to 000 00 automatically FREQ RESPONSE Menu Item This sets the frequency response step response and resolution of the channel accor...

Страница 64: ...l PEAK means the highest value of the channel VALLEY means the lowest value of the channel TRACK PEAK means the live tracking value and the highest value of the channel Only available on Model SC3004 TRACK VALLEY means the live tracking value and the lowest value of the channel Only available on Model SC3004 PEAK VALLEY means the highest and lowest values of the channel Only available on Model SC3...

Страница 65: ...rements or decrements This will make the display less sensitive to a noisy signal The choices are 00001 00002 00005 00010 00020 00050 00100 00200 This menu item may be automatically updated by a transducer s Signature Mod ule DISPLAY UNITS Menu Item Specifies the four character label that is displayed to the right of the channel s val ues When a character position is flashing press the UP or DOWN ...

Страница 66: ...ABLED means that activating the pin does nothing TRACK HOLD means that the tracking peak and valley values will not be updated HIGH LOW HOLD means that the peak and valley values will not be updated HIGH LOW CLEAR means that the peak and valley values are reset TARE ON activates the channel s Tare function resets it to zero TARE OFF deactivates the channel s Tare function As the Auxiliary Function...

Страница 67: ...tered in the ZERO SCALE VALUE register Next the instrument inserts Table 4 4 Comparison of Calibration Types Shunt Calibration mV V Calibration 2 3 or 5 point KnownLoad Calibration RECOMMENDED for most applications no yes no when frequently swapping transducers yes yes no when best possible accuracy required no no yes PROS and CONS Relative accuracy good better best Requires actual calibrated load...

Страница 68: ...he usual loads used are zero scale and full scale TYPE 3 POINT CAL means 3 Point Known Load Calibration You are prompted to apply the loads to the transducer that were entered in the KNOWN POINT 1 3 and KNOWN POINT 2 3 and KNOWN POINT 3 3 registers This technique can be used to compensate for the non linearity in the transducer The usual loads used are zero scale half scale and full scale but you ...

Страница 69: ...tem The transducer s zero scale output in engineering units which is usually zero This menu item may be automatically updated by a transducer s Signature Mod ule It is available only with the Shunt Calibration and mV V Calibration types SHUNT CAL VALUE Menu Item The shunt calibration value must be calculated by using information found on the transducer s Certificate of Calibration and the followin...

Страница 70: ...WN POINT 1 3 point 1 of 3 usually 0 of the transducer s full scale capacity KNOWN POINT 2 3 point 2 of 3 usually 50 of the transducer s full scale capacity KNOWN POINT 3 3 point 3 of 3 usually 100 of the transducer s full scale capacity When using the 5 Point Known Load Calibration type the following menu items are available KNOWN POINT 1 5 point 1 of 5 usually 0 of the transducer s full scale cap...

Страница 71: ...e previously entered as KNOWN POINT 1 2 and DISPLAY UNITS When you have applied this load to the transducer press ENTER The display will read WORKING then APPLY 01000 UNIT where 01000 and UNIT are previously entered as KNOWN POINT 2 2 and DISPLAY UNITS When you have applied this load to the transducer press ENTER Upon completion the display will indicate DONE and the instrument will return to the ...

Страница 72: ...AY UNITS When you have applied this load to the transducer press ENTER The display will read WORKING then APPLY 00750 UNIT where 00750 and UNIT are pre viously entered as KNOWN POINT 4 5 and DISPLAY UNITS When you have applied this load to the transducer press ENTER The display will read WORKING then APPLY 01000 UNIT where 01000 and UNIT are pre viously entered as KNOWN POINT 5 5 and DISPLAY UNITS...

Страница 73: ...n channel s tracking value the Analog Output will have the frequency response specified by the channel s FREQ RESPONSE menu item Otherwise the Analog Output will operate more slowly DAC ZERO SCALE Menu Item This specifies what value in engineering units corresponds to zero output on the Analog Output Zero output might be 0 Volts 2 5 Volts 5 Volts 4 mA or 12 mA depending on if the channel has a vol...

Страница 74: ...ions for this menu item are NO will not update the Signature Module YES will display WORKING update the Signature Module then display DONE DEFAULT SIG MOD Menu Item This menu item will overwrite any user settings in the Signature Module with the origi nal default settings The Signature Module always maintains a copy of the original default settings that were created at the factory The options for ...

Страница 75: ...K track ing value HI peak value and LO valley value when the VALUE button is pressed on the front panel The Analog to Digital converter counts are displayed as a percentage from 100 00 to 100 00 This can be used to verify that the transducer s output does not exceed the input range of the instrument If the Analog to Digital converter readings display more than 10 when there is no load on the trans...

Страница 76: ...nu Items The Analog Output can be driven by any channel s track peak or valley value See the Channel Menu section earlier in this chapter for a complete listing of SETUP menu items available on the DAC SETUP sub menu 10 6 3 Output Selection Jumpers located on the channel s circuit board determine what outputs are gener ated when the value selected to drive the Analog Output from the DAC CHANNEL an...

Страница 77: ...nel is equipped with see Identifying the Output Type on page 76 Use the DIAGNOSTICS DAC FULL SCALE and DIAGNOSTICS DAC ZERO SCALE menu items to force the Analog Output to a known output Then adjust your readout device panel meter PLC or data acquisition system to match Check the output selection jumpers see Output Selection on page 76 APPLY 00000 on Power up The channel has detected that the trans...

Страница 78: ...w on the rear panel of the instrument See Unamplified Transducer Connection to Strain Gage Input Channel on page 60 External Arrangement of AC powered SC1000 and SC2000 on page 27 and External Arrangement of Model SC3004 on page 33 Shunt Calibration Feature not Operating To activate the channel s shunt calibration function while the instrument is in RUN Mode see Shunt Calibration Check on page 21 ...

Страница 79: ...res adjustable digital low pass filtering 12 to 18 bit resolution depending on the filter setting and has several different input ranges These many input ranges allow 25 000 count resolution at the slowest filter setting for a wide variety of LVDT outputs The AC AC LVDT Input channel is calibrated to the transducer by using known displacement calibration Two rear panel control inputs can be field ...

Страница 80: ...g Return pins are electrically isolated from all other pins on the instrument Figure 11 1 Connection of Four or Five wire AC AC LVDT Table 5 5 AC AC LVDT Input Channel Pin Connections Pin Label Function Reference Pin 1 top EXC Excitation 10 2 N C No connection 3 N C No connection 4 EXC Excitation 10 5 SIG Signal 10 6 SIG Signal 10 7 OUT Analog Output 8 8 OUT Analog Return 9 N C No connection 10 DG...

Страница 81: ... menu item allows you to specify how many known displacements will be applied during calibration Step 3 Enter the CALIBRATION DATA Consult the Certificate of Calibration for the transducer when entering infor mation in the CALIBRATION DATA sub menu Step 4 Determine the LVDT s electrical null point then physically mount it The electrical null is the armature position that produces the least electri...

Страница 82: ...ow pass filter digital 24 tap FIR Resolution see page 83 Frequency Response see page 83 AUXILLIARY INPUTS Quantity 2 Type momentary contact closure Response Time 5ms Field Selectable Functions tare on tare off peak valley clear peak valley hold track hold ANALOG OUTPUT Voltage range voltage output channels 5 5 10 or 10 VDC Current range current output channels 4 20 mA Source any channel s track pe...

Страница 83: ...hifts in the transducer s output The two options for this menu item are ON and OFF FREQ RESPONSE Menu Item This sets the frequency response step response and resolution of the channel according to the table below The tracking value peak valley detector values and analog output on this channel are all affected The default value of 016 HERTZ is suitable for most applications NOTICE To compensate for...

Страница 84: ...l PEAK means the highest value of the channel VALLEY means the lowest value of the channel TRACK PEAK means the live tracking value and the highest value of the channel Only available on Model SC3004 TRACK VALLEY means the live tracking value and the lowest value of the channel Only available on Model SC3004 PEAK VALLEY means the highest and lowest values of the channel Only available on Model SC3...

Страница 85: ...l that is displayed to the right of the channel s val ues When a character position is flashing press the UP or DOWN button to change the character Press ENTER to advance to the next character DISPLAY AVERAGE Menu Item Controls the speed with which the channel s display values will update Display averaging does not affect the channel s analog output or its peak valley detection which will proceed ...

Страница 86: ...LEAR means that the peak and valley values are reset TARE ON activates the channel s Tare function resets it to zero TARE OFF deactivates the channel s Tare function As the Auxiliary Function pins are not isolated it is recommended that a push button switch or relay is used to connect these pins to the DGND pin ...

Страница 87: ...ique can be used to compensate for the different sensitivities that an LVDT may have on each side of its electrical null The usual displacements used are zero scale half scale and full scale but you are not restricted to these displacements TYPE 5 POINT CAL means 5 Point Known Displacement Calibration You are prompted to apply the displacements to the transducer that were entered in the KNOWN POIN...

Страница 88: ...er s full scale capacity When using the 3 Point Known Displacement Calibration type the following menu items are available KNOWN POINT 1 3 point 1 of 3 usually 0 of the transducer s full scale capacity KNOWN POINT 2 3 point 2 of 3 usually 50 of the transducer s full scale capacity KNOWN POINT 3 3 point 3 of 3 usually 100 of the transducer s full scale capacity When using the 5 Point Known Displace...

Страница 89: ...sly entered as KNOWN POINT 2 3 and DISPLAY UNITS When you have applied this displacement to the transducer press ENTER The display will read WORKING then APPLY 01000 UNIT where 01000 and UNIT are previously entered as KNOWN POINT 2 3 and DISPLAY UNITS When you have applied this displacement to the transducer press ENTER Upon completion the display will indicate DONE and the instrument will return ...

Страница 90: ...pre viously entered as KNOWN POINT 4 5 and DISPLAY UNITS When you have applied this displacement to the transducer press ENTER The display will read WORKING then APPLY 01000 UNIT where 01000 and UNIT are pre viously entered as KNOWN POINT 5 5 and DISPLAY UNITS When you have applied this displacement to the transducer press ENTER Upon completion the display will indicate DONE and the instrument wil...

Страница 91: ...ts lowest value VALLEY The options for this menu item are TRACK means the live tracking value of the channel PEAK means the highest value of the channel since the peak valley detector was last cleared VALLEY means the lowest value of the channel since the peak valley detector was last cleared If the DAC is set up to be driven by its own channel s tracking value the Analog Output will have the freq...

Страница 92: ...ommended option ON will allow the AD percentage of the Analog to Digital converter s full scale digitizing capability display source to be available along with TK tracking value HI peak value and LO valley value when the VALUE button is pressed on the front panel The Analog to Digital converter counts are displayed as a percentage from 100 00 to 100 00 This can be used to establish the transducer ...

Страница 93: ...o every calibration 11 6 2 Procedure Step 1 Use the channel s SETUP menu and set DIAGNOSTICS DISPLAY ADC to ON Exit the SETUP menu and re start the instrument Step 2 Display the channel to which the LVDT is connected Press and release the VALUE button until the AD display source is shown This is the raw analog to digital converter readings displayed as a percentage of its full scale 100 to 100 Ste...

Страница 94: ...nu Items The Analog Output can be driven by any channel s track peak or valley value See the Channel Menu section earlier in this chapter for a complete listing of SETUP menu items available on the DAC SETUP sub menu 11 7 3 Output Selection Jumpers located on the channel s circuit board determine what outputs are generated when the value selected to drive the Analog Output from the DAC CHANNEL and...

Страница 95: ...ent that the channel is equipped with see Identifying the Output Type on page 94 Use the DIAGNOSTICS DAC FULL SCALE and DIAGNOSTICS DAC ZERO SCALE menu items to force the Analog Output to a known output Then adjust your readout device panel meter PLC or data acquisition system to match Check the output selection jumpers see Output Selection on page 94 Auxiliary Function Pins Not Operating Make sur...

Страница 96: ...page 96 008 0608 00 ...

Страница 97: ...g 12 to 18 bit resolution depending on the filter setting and has several different input ranges These many ranges allow 50 000 count resolution at the slowest filter setting across a wide variety of input ranges A combination of SETUP menu items and hardware jumpers are used to config ure the excitation supply output voltage 28 VDC 15 VDC 15 VDC or 12 VDC the input type voltage or current and inp...

Страница 98: ...ns Pin Label Function Reference Pin 1 top EXC Excitation 10 2 SHUNT1 Shunt Cal Relay 3 3 SHUNT2 Shunt Cal Relay 2 4 EXC Excitation 10 5 SIG Signal 10 6 SIG Signal 10 7 OUT Analog Output 8 8 OUT Analog Return 9 N C No Connection 10 DGND Digital Ground 11 AUX1 Auxiliary Function 1 connect to pin 10 to activate 10 12 bottom AUX2 Auxiliary Function 2 connect to pin 10 to activate 10 CAUTION Identify t...

Страница 99: ...iversal In Line amplifiers with shunt cal Models JEC replaces model MDL JEC AG replaces model DLA JEC C replaces model MDLC DW7U DW7C and DW7S DC DC LVDTs no shunt cal The High Level Input channel s Configuration Jumpers must be set as follows for proper operation See Excitation and Signal Jumpers on page 106 Excitation supply 15 VDC Excitation supply 15 VDC Signal type voltage Signal reference si...

Страница 100: ...ransducers with Option 2c or Option 2t internal amplifiers with shunt cal Models UV or UV 10 Universal In Line amplifiers with shunt cal The High Level Input channel s Configuration Jumpers must be set as follows for proper operation See Excitation and Signal Jumpers on page 106 Excitation supply 28 VDC Excitation supply GND Signal type voltage Signal reference differential Figure 12 2 3 wire Volt...

Страница 101: ... High Level Input channel An example of a such device includes transducers with Option 2d or Option 2g internal amplifiers with shunt cal The High Level Input channel s Configuration Jumpers must be set as follows for proper operation See Excitation and Signal Jumpers on page 106 Excitation supply 28 VDC Excitation supply GND Signal type voltage Signal reference differential Figure 12 3 3 wire Vol...

Страница 102: ...include transducers with the Option 2j internal amplifier with shunt cal Model U3W Universal In Line amplifiers with shunt cal The High Level Input channel s Configuration Jumpers must be set as follows for proper operation See Excitation and Signal Jumpers on page 106 Excitation supply 28 VDC Excitation supply GND Signal type current Signal reference single ended Figure 12 4 3 wire Current Amp Co...

Страница 103: ... the Option 2k internal amplifier not equipped with shunt cal transducers with the Option 2k 3d internal amplifier buffered shunt cal Model U2W Universal In Line amplifiers not equipped with shunt cal The High Level Input channel s Configuration Jumpers must be set as follows for proper operation See Excitation and Signal Jumpers on page 106 Excitation supply 28 VDC Excitation supply GND Signal ty...

Страница 104: ...cers with the Option 2p 3d internal amplifier single wire shunt cal transducers with the Option 2y internal amplifier not equipped with shunt cal transducers with the Option 2y 3d internal amplifier single wire shunt cal The High Level Input channel s Configuration Jumpers must be set as follows for proper operation See Excitation and Signal Jumpers on page 106 Excitation supply 28 VDC Excitation ...

Страница 105: ...ut channel Examples of such devices include Models DLB DLE and DLF DC DC LVDTs The High Level Input channel s Configuration Jumpers must be set as follows for proper operation See Excitation and Signal Jumpers on page 106 Excitation supply 12 VDC Excitation supply GND Signal type voltage Signal reference single ended Figure 12 7 Low Voltage DC DC LVDT Connection to High Level Input Channel ...

Страница 106: ...he channel board from the chassis Step 3 Change the jumper settings according to the figure below Step 4 Re install the channel board and replace the rear panel Figure 12 8 Excitation and Signal Jumper Locations on the High Level Input Channel CAUTION Wiring on page 98 explains which jumpers settings are required for a particu lar transducer type Incorrect placement of the Excitation and Signal ju...

Страница 107: ...age 98 and Excitation and Signal Jumpers on page 106 Step 3 Enter the CALIBRATION TYPE There are two methods that can be used to calibrate the transducer to the Input Channel Each has advantages and disadvantages as described in CALIBRATION TYPE Menu Item on page 113 It is important to know your application in order to select the appropriate calibration type Step 4 Enter the CALIBRATION DATA Other...

Страница 108: ...er 24 bit Sigma Delta Low pass filter digital 24 tap FIR Resolution and Frequency Response see FREQ RESPONSE Menu Item on page 109 INSTRUMENT ONLY ACCURACY Frequency Response setting 16Hz Linearity repeat ability hysteresis Known Load Calibration 0 01 F S AUXILLIARY INPUTS Quantity 2 Type momentary contact closure Response Time 5ms Field Selectable Functions tare on tare off peak valley clear peak...

Страница 109: ...ts If the tracking value never leaves the range of 000 10 to 000 10 for any 10 second period then tracking value will reset to 000 00 automatically FREQ RESPONSE Menu Item This sets the frequency response step response and resolution of the channel according to the table below The tracking value peak valley detector values and analog output on this channel are all affected The default value of 016...

Страница 110: ...TRACK PEAK means the live tracking value and the highest value of the channel Only available on Model SC3004 TRACK VALLEY means the live tracking value and the lowest value of the channel Only available on Model SC3004 PEAK VALLEY means the highest and lowest values of the channel Only available on Model SC3004 ...

Страница 111: ...itive to a noisy signal The choices are 00001 00002 00005 00010 00020 00050 00100 00200 DISPLAY UNITS Menu Item Specifies the four character label that is displayed to the right of the channel s val ues This menu item doesn t change the mathematical scaling of the channel s values that can be changed by altering the CALIBRATION DATA FULL SCALE VALUE menu item and then performing a re calibration W...

Страница 112: ... activating the pin does nothing TRACK HOLD means that the tracking peak and valley values will not be updated HIGH LOW HOLD means that the peak and valley values will not be updated HIGH LOW CLEAR means that the peak and valley values are reset TARE ON activates the channel s Tare function resets it to zero TARE OFF deactivates the channel s Tare function As the Auxiliary Function pins are not is...

Страница 113: ...trument then takes a reading and adjusts itself using the SHUNT CAL VALUE register TYPE 2 POINT CAL means 2 Point Known Load Calibration You are prompted to apply the loads to the transducer that were entered in the KNOWN POINT 1 2 and KNOWN POINT 2 2 registers This technique assumes that the transducer is linear so the usual loads used are zero scale and full scale TYPE 3 POINT CAL means 3 Point ...

Страница 114: ...loads to the transducer that were entered in the KNOWN POINT 1 5 KNOWN POINT 2 5 KNOWN POINT 3 5 KNOWN POINT 4 5 and KNOWN POINT 5 5 regis ters This technique can be used to compensate for the non linearity in the transducer The usual loads used are zero scale one quarter scale half scale three quarters scale and full scale but you are not restricted to these loads The selection of the calibration...

Страница 115: ...ansducer s full scale output in Volts or milliamps Sometimes called Calibration Factor Full Scale Capacity The transducer s desired full scale reading in engineering units PSI lbs Kg etc You may wish to convert the engineering units of this value For example if the transducer full scale is 3000 PSIG and you wish the channel to read in bar you must convert 3000 PSIG into 206 84 bar This menu item i...

Страница 116: ...POINT 3 5 point 3 of 5 usually 50 of the transducer s full scale capac ity KNOWN POINT 4 5 point 4 of 5 usually 75 of the transducer s full scale capac ity KNOWN POINT 5 5 point 5 of 5 usually 100 of the transducer s full scale capacity These menu items are only available with the Known Load Calibration types NOTICE To insure both correct operation of the transducer and application of the load the...

Страница 117: ...TER Upon completion the display will indicate DONE and the instrument will return to the RUN mode If the CALIBRATION TYPE is 3 Point Known Load Calibration The display will read DOING 3POINT CAL and prompt you to APPLY 00000 UNIT where 00000 and UNIT are previously entered as KNOWN POINT 1 3 and DISPLAY UNITS When you have applied this load to the transducer press ENTER The display will read WORKI...

Страница 118: ...The display will read WORKING then APPLY 00750 UNIT where 00750 and UNIT are previously entered as KNOWN POINT 4 5 and DISPLAY UNITS When you have applied this load to the transducer press ENTER The display will read WORKING then APPLY 01000 UNIT where 01000 and UNIT are previously entered as KNOWN POINT 5 5 and DISPLAY UNITS When you have applied this load to the transducer press ENTER The displa...

Страница 119: ...em are TRACK means the live tracking value of the channel PEAK means the highest value of the channel since the peak valley detector was last cleared VALLEY means the lowest value of the channel since the peak valley detector was last cleared If the DAC is set up to be driven by its own channel s tracking value the Analog Output will have the frequency response specified by the channel s FREQ RESP...

Страница 120: ...ale digitizing capability display source to be available along with TK track ing value HI peak value and LO valley value when the VALUE button is pressed on the front panel The Analog to Digital converter counts are displayed as a percentage from 100 00 to 100 00 This can be used to verify that the transducer s output does not exceed the input range of the instrument If the Analog to Digital conve...

Страница 121: ...elow 12 7 2 Channel Menu Items The Analog Output can be driven by any channel s track peak or valley value See the Channel Menu section earlier in this chapter for a complete listing of SETUP menu items available on the DAC SETUP sub menu 12 7 3 Output Selection Jumpers located on the channel s circuit board determine what outputs are gener ated when the value selected to drive the Analog Output f...

Страница 122: ...g the transducer may have a high zero offset Use the channel s SETUP menu and set DIAGNOSTICS DISPLAY ADC to ON in the RUN mode this will allow the VALUE but ton to display raw A D readings as a percentage of its full scale If the raw A D readings display more than 10 when there is no load on the transducer the transducer has a high zero offset Analog Output Incorrect Make certain of the type of A...

Страница 123: ...t 5 Limit 6 Limit 7 and Limit 8 Its relay outputs show the status of these limits In the SYSTEM CONFIGURATION sub menu channels of this type are identified as LIMIT 05 08 13 1 3 Third Channel Installed The third Relay Output channel installed in an instrument adds another four limits to the instrument Limit 9 Limit 10 Limit 11 and Limit 12 Its relay outputs show the status of these limits In the S...

Страница 124: ...p RELAY1 NC Relay 1 normally closed 2 2 RELAY1 C Relay 1 common 3 RELAY1 NO Relay 1 normally open 2 4 RELAY2 NC Relay 2 normally closed 5 5 RELAY2 C Relay 2 common 6 RELAY2 NO Relay 2 normally open 5 7 RELAY3 NC Relay 3 normally closed 8 8 RELAY3 C Relay 3 common 9 RELAY3 NO Relay 3 normally open 8 10 RELAY4 NC Relay 4 normally closed 11 11 RELAY4 C Relay 4 common 12 bottom RELAY4 NO Relay 4 norma...

Страница 125: ...Menu Reference on page 157 Step 1 Wire to the channel s connector See the Wiring section earlier in this chapter for details Step 2 Use the appropriate Limit Menu to setup the limits See Limits on page 53 for information about limits NOTICE A SensoCode program running on a Mathematics Virtual Channel may over ride the default behavior of the relay outputs Consult the Customer Information Sheet inc...

Страница 126: ...page 126 008 0608 00 13 4 Specifications RELAY OUTPUT Quantity and Type 4 form C Contact Rating 0 5A 50VAC max consult factory for 125VAC max operation 1A 30VDC max ...

Страница 127: ... in SETUP Menu mode on page 23 A diagram of all menus is located in Setup Menu Reference on page 157 13 5 1 DIAGNOSTICS Sub Menu RELAY 1 Menu Item The options for this menu item are ON will turn the relay on OFF will turn the relay off RELAY 2 Menu Item The options for this menu item are ON will turn the relay on OFF will turn the relay off RELAY 3 Menu Item The options for this menu item are ON w...

Страница 128: ...page 128 008 0608 00 ...

Страница 129: ...14 DAC Output Channel 14 1 Features The DAC Output channel uses a digital to analog converter to generate a voltage or current from any channel s track peak or valley value This type of channel is often used to provide a Mathematics Virtual channel with an analog output ...

Страница 130: ...connector is shown on the following table The Analog Output and Analog Return pins are electrically isolated from all other pins on the instrument Table 3 9 DAC Output Channel Pin Connections Pin Label Function Reference Pin 1 top N C No connection 2 N C No connection 3 N C No connection 4 N C No connection 5 N C No connection 6 N C No connection 7 OUT Analog Output 8 8 OUT Analog Return 9 N C No ...

Страница 131: ...ut channel or a Current DAC Output channel Check the instrument s Customer Information Sheet or see Identifying the Output Type on page 135 Step 2 Wire the readout instrument to the channel s connector See the Wiring section earlier in this chapter for details Step 3 Select the value channel and source used to drive the Analog Output See the Channel Menu on page 133 for details Step 4 Select the f...

Страница 132: ...T Voltage range voltage output channels 5 5 or 10 VDC Current range current output channels 4 20 mA Source any channel s track peak or valley value Isolation 500V Resolution 12 bits Update Rate depends on how many channels are installed in the instrument ...

Страница 133: ...el PEAK means the highest value of the channel since the peak valley detector was last cleared VALLEY means the lowest value of the channel since the peak valley detector was last cleared DAC ZERO SCALE Menu Item This specifies what value in engineering units corresponds to zero output on the Analog Output Zero output might be 0 Volts 2 5 Volts 5 Volts 4 mA or 12 mA depending on if the channel has...

Страница 134: ... output then DAC UPDATED is displayed This is useful when calibrating or trimming the readout connected to the Analog Output DAC ZERO SCALE Menu Item When this menu item is selected the Analog Output of the channel is forced to its zero scale output then DAC UPDATED is displayed This is useful when calibrating or trim ming the readout connected to the Analog Output ...

Страница 135: ...ems The Analog Output can be driven by any channel s track peak or valley value See the Channel Menu section earlier in this chapter for a complete listing of SETUP menu items available on the DAC SETUP sub menu 14 6 3 Output Selection Jumpers located on the channel s circuit board determine what outputs are gener ated when the value selected to drive the Analog Output from the DAC CHANNEL and DAC...

Страница 136: ...d DIAGNOSTICS DAC ZERO SCALE menu items to force the Analog Output to a known output Then adjust your readout device panel meter PLC or data acquisition system to match Check the output selection jumpers see Output Selection on page 135 Check the DAC SETUP DAC ZERO SCALE and DAC SETUP DAC FULL SCALE menu items Sensitivity to EMI RFI To obtain maximum immunity to electromagnetic or radio frequency ...

Страница 137: ...SC Series Instruction Manual page 137 DAC Output Channel 14 ...

Страница 138: ...page 138 008 0608 00 ...

Страница 139: ...ver already allows an Input Channel to show its own Track Peak Track Valley or Peak Valley values on the display at the same time Thus a Split Display Virtual Channel is not generally used on an Model SC3004 15 2 Wiring A Virtual Channel exists in software only it does not occupy a physical slot inside of the instrument s chassis However a Virtual Channel does require a channel number There is no ...

Страница 140: ...l that is displayed on the right half of the split display RIGHT SOURCE Menu Item This lets you designate the data source of the channel monitored by the right half of the split display Each channel has three data sources the live tracking value TRACK its highest value PEAK and its lowest value VALLEY The options for this menu item are TRACK means the live tracking value of the channel PEAK means ...

Страница 141: ...SC Series Instruction Manual page 141 Split Display Virtual Channel 15 ...

Страница 142: ...page 142 008 0608 00 ...

Страница 143: ...nso Code programs are created These SensoCode programs are loaded into the Mathematics Virtual channels of the instrument at the factory The end result is an instrument that meets your specific needs but is built with off the shelf hardware and software SensoCode programs allow the instrument to perform a variety of functions including Mathematics addition subtraction multiplication division squar...

Страница 144: ...is However a Virtual Channel does require a channel number A SensoCode program that is executed by a Mathematics Virtual Channel can use the System connector s Limit Output pins and Function Input pins for its own purposes Consult the Customer Information Sheet and Supplemental Instructions included with the particular instrument for details ...

Страница 145: ...Reference on page 157 Step 1 Examine the Supplemental Instructions Supplemental Instructions contain important information about what each Mathematics Channel does and what Function Inputs and or Limit Output pins of the System connector are used Step 2 Wire to the pins on the System connector See System Connector on page 35 It is important to read the Supple mental Instructions to determine the w...

Страница 146: ...ER VALUE 3 and USER VALUE 4 However a SensoCode program can alter the menu item text for each of the User Values to make them more relevant to the operator For example USER VALUE 1 might be re titled JELLY BEAN COUNT 16 4 3 VIEW SENSOCODE Menu Item This menu item allows the display of each step either an operand or an instruction of the SensoCode program presently loaded into a Mathematics Virtual...

Страница 147: ...to move the decimal point to the right and left DISPLAY COUNT BY Menu Item Determines by what value the display increments or decrements This will make the display less sensitive to a noisy signal The choices are 00001 00002 00005 00010 00020 00050 00100 00200 DISPLAY UNITS Menu Item Specifies the four character label that is displayed to the right of the channel s val ues This menu item doesn t c...

Страница 148: ...lue of the channel VALLEY means the lowest value of the channel For multi line displays the choices are TRACK means the live tracking value of the channel PEAK means the highest value of the channel VALLEY means the lowest value of the channel TRACK PEAK means the live tracking value and the highest value of the channel TRACK VALLEY means the live tracking value and the lowest value of the channel...

Страница 149: ...Messages on page 151 for information relating to error messages NOTICE Supplemental Instructions included with the instrument may contain important information about which indicator lights Function Input pins and or Limit Output pins of the System connector are used by the Mathematics Virtual channel ...

Страница 150: ...page 150 008 0608 00 ...

Страница 151: ... modes it stops operation and enters its ERROR mode as described in Operating Modes on page 19 The most frequent causes of error messages are Wiring electrical connection problems Misplaced address jumpers when installing or removing channel circuit boards Swapping channel circuit boards between instruments Incorrect calibration of the transducer to the instrument ...

Страница 152: ...oftware stack has overflowed Restart the instrument If this error persists contact our Service department with the information found in the SYSTEM MENU SOFTWARE REV menu item Error 22 TYPE NOT FOUND The instrument has not been programmed to understand the type of channel Contact our Service department with the information found in the SYSTEM MENU SOFTWARE REV menu item Error 23 CAN T RENDER FS The...

Страница 153: ...AD PERSONALITY The channel doesn t understand the given card type Contact Sensotec with the information found in the SYSTEM MENU SOFTWARE REV menu item Error 38 BAD FORMAT The channel s EEPROM memory contains a display setting which is not under stood by the instrument Check that every item in the channel s DISPLAY SETUP menu has a valid setting Error 39 WRONG HARDWARE The channel type expected by...

Страница 154: ...was found inside a Signature Module during a read opera tion Check the wiring Error 54 SIG MOD LENGTH W An unexpected record type was found inside a Signature Module during a write opera tion Check the wiring and re start the instrument Error 55 SIG MOD VERIFY SP A data verification error occurred during a write operation to a Signature Module s scratchpad Check the wiring and re start the instrum...

Страница 155: ...alibration or mV V Calibration However since the channel s CALIBRATION TYPE menu item is set to one of the Known Load Calibration settings no calibration can take place Either re connect the original transducer or use the SETUP menu to perform a calibration Error 70 CAN T DO GAIN The channel s EEPROM memory contains an input range setting which is beyond the capabilities of the channel Change the ...

Страница 156: ... Instruction is beyond the end of the SensoCode pro gram The SensoCode program loaded in the Mathematics Virtual channel is invalid Error 87 GOTO NOT JUMP The SensoCode program contains a Go to Instruction that has not been converted into a Jump Instruction Restart the instrument ...

Страница 157: ...Setup Menu Reference 18 Chapter 18 Setup Menu Reference 18 1 Navigation instructions Press UP DOWN to enter the SETUP menu mode Press UP to move up Press DOWN to move down Press ENTER to select an item Press EXIT to re start the instrument ...

Страница 158: ...varies according to channel type CHANNEL 17 MENU menu varies according to channel type CHANNEL 18 MENU menu varies according to channel type CHANNEL 19 MENU menu varies according to channel type CHANNEL 20 MENU menu varies according to channel type CHANNEL 21 MENU menu varies according to channel type CHANNEL 22 MENU menu varies according to channel type CHANNEL 23 MENU menu varies according to ch...

Страница 159: ...TYPE CHANNEL 16 TYPE CHANNEL 17 TYPE CHANNEL 18 TYPE CHANNEL 19 TYPE CHANNEL 20 TYPE CHANNEL 21 TYPE CHANNEL 22 TYPE CHANNEL 23 TYPE LEAVE SUB MENU DIAGNOSTICS OUTPUT 1 PIN 14 ON page 41 OFF OUTPUT 2 PIN 15 ON OFF OUTPUT 3 PIN 16 ON OFF OUTPUT 4 PIN 17 ON OFF INPUT TEST page 41 SCAN TIME page 41 LEAVE SUB MENU INSTALL CHANNEL 65 STRAIN GAGE V page 42 66 STRAIN GAGE I 67 HIGH LEVEL V 68 HIGH LEVEL ...

Страница 160: ... TRACK page 57 PEAK VALLEY LEAVE MENU Table 7 13 SERIAL COMMUNICATIONS Menu SERIAL COM MENU INTERFACE page 49 ADDRESS page 49 BAUD RATE page 49 AUTO LINE FEED page 49 TRANSMIT TEST page 49 LEAVE MENU Table 7 14 DISPLAY Menu for Models SC2000 SC2001 DISPLAY MENU UPPER CHANNEL page 51 LOWER CHANNEL page 51 LOWER MODE BLANK page 51 LIMITS 01 04 CHANNEL DISPLAY DISABLE ON page 51 OFF LEAVE MENU Table ...

Страница 161: ...E SUB MENU DISPLAY SETUP DISPLAY DIGITS 5 DIGIT BIPOLAR page 65 6 DIGIT UNIPOLAR 7 DIGIT UNIPOLAR DISPLAY DECPT 99999 page 65 9999 9 999 99 99 999 9 9999 99999 DISPLAY COUNTBY 00001 page 65 00002 00005 00010 00020 00050 00100 00200 DISPLAY UNITS page 65 DISPLAY AVERAGE ON page 65 OFF LEAVE SUB MENU AUX1 FUNCTION DISABLED page 59 TRACK HOLD HIGH LOW HOLD HIGH LOW CLEAR TARE ON TARE OFF AUX2 FUNCTIO...

Страница 162: ...D POINT 4 5 KNOWN LOAD POINT 5 5 LEAVE SUB MENU CALIBRATE page 71 DAC SETUP DAC CHANNEL page 73 DAC SOURCE TRACK PEAK VALLEY DAC ZERO SCALE DAC FULL SCALE LEAVE SUB MENU SIGNATURE MODULE UPDATE SIG MOD NO page 74 YES DEFAULT SIG MOD NO YES UPDATE DAC SETUP NO YES LEAVE SUB MENU DIAGNOSTICS DAC FULL SCALE page 75 DAC ZERO SCALE VERSION INFO DISPLAY ADC ON OFF LINEARIZATION ON OFF DISABLE CHANNEL ON...

Страница 163: ...UB MENU DISPLAY SETUP DISPLAY DIGITS 5 DIGIT BIPOLAR page 85 6 DIGIT UNIPOLAR 7 DIGIT UNIPOLAR DISPLAY DECPT 99999 page 85 9999 9 999 99 99 999 9 9999 99999 DISPLAY COUNTBY 00001 page 85 00002 00005 00010 00020 00050 00100 00200 DISPLAY UNITS page 85 DISPLAY AVERAGE ON page 85 OFF LEAVE SUB MENU AUX1 FUNCTION DISABLED page 85 TRACK HOLD HIGH LOW HOLD HIGH LOW CLEAR TARE ON TARE OFF AUX2 FUNCTION D...

Страница 164: ...OAD POINT 4 5 KNOWN LOAD POINT 5 5 LEAVE SUB MENU CALIBRATE page 89 DAC SETUP DAC CHANNEL page 91 DAC SOURCE TRACK PEAK VALLEY DAC ZERO SCALE DAC FULL SCALE LEAVE SUB MENU DIAGNOSTICS DAC FULL SCALE page 92 DAC ZERO SCALE VERSION INFO DISPLAY ADC ON OFF LINEARIZATION ON OFF LEAVE SUB MENU LEAVE MENU Table 7 17 AC AC LVDT Input Channel Menu Continued ...

Страница 165: ...B MENU DISPLAY SETUP DISPLAY DIGITS 5 DIGIT BIPOLAR page 111 6 DIGIT UNIPOLAR 7 DIGIT UNIPOLAR DISPLAY DECPT 99999 page 111 9999 9 999 99 99 999 9 9999 99999 DISPLAY COUNTBY 00001 page 111 00002 00005 00010 00020 00050 00100 00200 DISPLAY UNITS page 111 DISPLAY AVERAGE ON page 111 OFF LEAVE SUB MENU AUX1 FUNCTION DISABLED page 112 TRACK HOLD HIGH LOW HOLD HIGH LOW CLEAR TARE ON TARE OFF AUX2 FUNCT...

Страница 166: ...D POINT 2 5 KNOWN LOAD POINT 3 5 KNOWN LOAD POINT 4 5 KNOWN LOAD POINT 5 5 LEAVE SUB MENU CALIBRATE page 117 DAC SETUP DAC CHANNEL page 119 DAC SOURCE TRACK PEAK VALLEY DAC ZERO SCALE DAC FULL SCALE LEAVE SUB MENU DIAGNOSTICS DAC FULL SCALE page 120 DAC ZERO SCALE VERSION INFO DISPLAY ADC ON OFF LINEARIZATION ON OFF DISABLE CHANNEL ON OFF LEAVE SUB MENU LEAVE MENU Table 7 18 High Level Input Chann...

Страница 167: ...LEAVE MENU Table 7 20 DAC Output Channel Menu CHANNEL xx MENU DAC SETUP DAC CHANNEL page 133 DAC SOURCE TRACK PEAK VALLEY DAC ZERO SCALE DAC FULL SCALE LEAVE SUB MENU DIAGNOSTICS DAC FULL SCALE page 134 DAC ZERO SCALE LEAVE SUB MENU LEAVE MENU Table 7 21 Split Display Virtual Channel CHANNEL xx MENU LEFT CHANNEL page 140 LEFT SOURCE TRACK page 140 PEAK VALLEY RIGHT CHANNEL page 140 RIGHT SOURCE TR...

Страница 168: ...page 146 COMMAND 1 COMMAND 2 COMMAND 3 COMMAND 4 COMMAND 5 COMMAND 6 COMMAND 7 LEAVE SUB MENU DISPLAY SETUP DISPLAY DIGITS 5 DIGIT BIPOLAR page 147 6 DIGIT UNIPOLAR 7 DIGIT UNIPOLAR DISPLAY DECPT 99999 page 147 9999 9 999 99 99 999 9 9999 99999 DISPLAY COUNTBY 00001 page 147 00002 00005 00010 00020 00050 00100 00200 DISPLAY UNITS page 147 DISPLAY AVERAGE ON page 147 OFF LEAVE SUB MENU POWER ON SOU...

Страница 169: ... channel 85 95 High Level Input channel 112 122 Strain Gage Input channel 66 77 B buttons CHANNEL 21 CLEAR 20 57 DOWN 23 ENTER 23 EXIT 23 TARE 21 UP 23 VALUE 20 C calibration AC AC LVDT Input channel 81 High Level Input channel 107 Strain Gage Input channel 61 case removal of SC1000 and SC2000 28 of SC2001 31 of SC3004 34 channel address jumpers 42 changing which is displayed 21 51 clearing peak a...

Страница 170: ... 116 118 Analog Output 119 122 auto zero function 109 Auxilliary Function pins 112 122 calibration procedure 107 calibration types comparison 113 display setup 111 EMI RFI interference 122 frequency response 109 OVLD message 122 remote control of Track Hold Peak Valley clear Peak Valley hold or Tare 112 122 resolution 109 SETUP menu 109 Shunt Calibration 21 113 115 117 software revision 120 specif...

Страница 171: ...gation instructions 157 overview 23 reference 158 shunt calibration check 21 Signature Calibration defined 14 with Strain Gage Input channel 67 74 Signature Module 74 defaulting to factory settings 74 updating channel s Analog Ouput settings 74 writing with information from SETUP menus 74 software revision AC AC LVDT Input channel 92 chassis 41 High Level Input channel 120 Mathematics Virtual chan...

Страница 172: ...activating 21 remote of all channels 36 See also Auxilliary Function pins Tare light 21 troubleshooting AC AC LVDT Input channel 95 DAC Output channel 136 High Level Input channel 122 Strain Gage Input channel 77 U UP button 23 user values 146 V VALUE button 20 W wiring AC AC LVDT Input channel 80 DAC Output channel 130 High Level Input channel 98 Mathematics Virtual channel 144 Relay Output chann...

Страница 173: ...he foregoing is the Buyer s sole remedy and is in lieu of all other warranties expressed or implied including those of merchantability and fitness for a particular purpose Specifications may change without notice The information we supply is believed to be accurate and reliable as of this printing However we assume no responsi bility for its use While we provide applications assistance personally ...

Страница 174: ...A Copyright 2005 Honeywell International Inc Honeywell Automation and Control Solutions Sensotec Sensors 2080 Arlingate Lane Columbus Ohio 43228 4112 Tel 614 850 5000 Fax 614 850 1111 E mail service sensotec com http www honeywell com sensotec ...

Отзывы: