INTELLIDOX DOCKING MODULE USER MANUAL || REPLACE DETECTOR
CRADLE AND CALIBRATION INSERT
HONEYWELL
PAGE 95 OF 119
Replace the Calibration Insert
To replace the calibration insert, remove the calibration insert from the IntelliDoX (unscrew
the screws and detach the gas tubes), and then install the replacement calibration insert.
1. Disconnect the power cord, network cable, tubing, and inlet filter assemblies from the
IntelliDoX.
2. Use a screwdriver to unscrew the two screws from the calibration insert.
3. Detach the two gas tubes from the IntelliDoX.
4. Remove the calibration insert.
5. Get the replacement calibration insert.