Update Module Firmware
Honeywell is committed to continuously improving product features and performance. As a
result, firmware updates may become available for the modules. Honeywell recommends
that you visit the appropriate product site at
https://www.honeywellanalytics.com/en/products/IntelliDoX
latest firmware version for your modules.
Updating Module Firmware via USB Flash Drive
Follow these instructions to update firmware for an individual module or a group of up to
five connected modules. The update is processed only when the module or group of
modules is idle.
For connected modules, you must transfer the firmware update to each module
in the group.
1. Download and save the firmware update file to a PC or network drive. Do not
rename the file.
2. Transfer the update to a USB flash drive.
a) Insert a USB flash drive into a USB port on your PC. Use Windows
Explorer or My Computer to open the USB flash drive, and then
create a new folder named
System
.
b) Copy the firmware update to the
System
folder on the USB flash
drive. Do not change the file name or extension.
c) Eject the USB flash from the PC.
3. Transfer the file from the USB flash drive to a module.
a) Insert the USB flash drive into the USB port on the module. The
USB File Copy Utility
menu is displayed.
b) Press or to select
Copy to dock
and then press to
continue.
c) The LCD changes to yellow. A progress bar and alternating
Do not
remove stick
and
USB transfer in progress
messages are
displayed while the file is transferred.
d) After the transfer is successfully completed, the LCD changes to
blue and the
USB File Copy Utility
menu is displayed. Press or
to select
Exit
. Press to continue.
e) The LCD changes to green. Remove the USB flash drive from the
port when prompted.
4. The firmware update begins automatically when the module is idle. If the module is
part of a group of connected modules, then installation begins after all of the
modules in the group are idle.
5. The LCD changes to yellow.
Installing new dock firmware. Please wait …
is
displayed.
Do not deactivate the module during installation.
6. After the update is installed, updated modules restart automatically.
IntelliDoX
reboot pending. Please wait…
is displayed. After a few minutes, IntelliDoX
rebooting is displayed. The reboot takes about 5 seconds. The screen deactivates
briefly while the module restarts.
7. After the installation is complete, the LCD changes to green and
Firmware
update complete
is displayed.
8. Verify the version of the installed firmware.