8.5. Checking the functions before using the instrument
Before installing the HERMetic instrument as described in section 9.10, the following tests are recommended
to ensure that the instrument is ready to work.
8.5.1. Battery
Refer to section 10.2 "
8.5.2. Temperature
Switch on the unit.
The buzzer shall beep every 2 sec.
When the temperature is displayed, check that it
shows the surrounding temperature.
8.5.3. Ullage
Switch on the unit.
The buzzer shall beep every 2 sec.
Check the ullage in a glass of water.
Check the ullage by immersing the ultrasonic gap
sensor but not the electrodes (position A); The
buzzer shall beep continuously.
8.5.4. Interface
Switch on the unit.
The buzzer shall beep every 2 sec.
Check the interface in a glass of water.
Check the interface by immersing the interface
electrodes also (position B). The buzzer shall
beep intermittently.
position A
position B
Figure 9-7
Содержание HERMetic UTImeter Gtex Chem IIC
Страница 5: ...50489 GTEX Chem IIC 1805 5 UTImeter Gtex Chem IIC Figure 2 2 ...
Страница 15: ...50489 GTEX Chem IIC 1805 UTImeter Gtex Chem IIC 15 Figure 7 2 ...
Страница 55: ...50489 GTEX Chem IIC 1805 UTImeter Gtex Chem IIC 55 Figure 13 1 general assembly list of the main spare parts ...
Страница 56: ...50489 GTEX Chem IIC 1805 UTImeter Gtex Chem IIC 56 Figure 13 2 display unit assembly TS 10227 details ...
Страница 59: ...50489 GTEX Chem IIC 1805 UTImeter Gtex Chem IIC 59 Figure 13 5 storage tube SS1 Q1 TS 10186 details ...
Страница 60: ...50489 GTEX Chem IIC 1805 UTImeter Gtex Chem IIC 60 Figure 13 6 storage tube SS1 Q2 TS 10187 details ...
Страница 63: ......
Страница 64: ......
Страница 65: ......
Страница 66: ......
Страница 67: ......
Страница 68: ......
Страница 69: ......
Страница 71: ......
Страница 72: ......
Страница 73: ......
Страница 75: ......
Страница 76: ......
Страница 77: ......
Страница 78: ......
Страница 79: ......
Страница 80: ......