MAINTENANCE, CLEANING AND REPAIR
Information for general use
Rev. M / 73023639
238
GasLab Q2
10.1.10.
System tightness test
Leaks in the gas supply lines increase the danger of explosion and may result in
inaccurate measurement results.
Les fuites dans les conduites
d’alimentation en gaz augmentent le risque d’explosion et
peuvent entraîner des mesures défaillantes!
Step
Action: Tightness test
1.
If the device is already operating, close the valves in the GasLab Q2 by switching off
the power supply.
2.
Supply all inlet gas lines for the measuring equipment with gases and pressures so that
the pipelines contain a gas volume.
3.
Check all the parts of the system for leaks using a gas detector.
4.
If you identify a leak, it must be sealed immediately. Shut off the gas supply. Seal
the leak. Repeat the test.
5.
Do not complete any further action or work
until the system is completely tight.